The purpose was to compare observer performance in the detection of breast cancer using hard-copy film, and 3-megapixel (3-MP) and 5-megapixel (5-MP) liquid crystal display (LCD) monitors in a simulated screening setting. We amassed 100 sample sets, including 32 patients with surgically proven breast cancer (masses present, N = 12; microcalcifications, N = 10; other types, N = 10) and 68 normal controls. All the mammograms were obtained using computed radiography (CR; sampling pitch of 50 mum). Twelve mammographers independently assessed CR mammograms presented in random order for hard-copy and soft-copy reading at minimal 4-week intervals. Observers rated the images on seven-point (1 to 7) and continuous (0 to 100) malignancy scales. Receiver-operating-characteristics analysis was performed, and the average area under the curve (AUC) was calculated for each modality. The jackknife method with the Bonferroni correction was applied to multireader/multicase analysis. The average AUC values for the 3-MP LCD, 5-MP LCD, and hard-copy film were 0.954, 0.947, and 0.956 on the seven-point scale and 0.943, 0.923, and 0.944 on the continuous scale, respectively. There were no significant differences among the three modalities on either scale. Soft-copy reading using 3-MP and 5-MP LCDs is comparable to hard-copy reading for detecting breast cancer.
目的是在模拟筛查环境中比较使用硬拷贝胶片、300万像素(3-MP)和500万像素(5-MP)液晶显示器(LCD)检测乳腺癌时观察者的表现。我们收集了100个样本集,包括32例经手术证实患有乳腺癌的患者(有肿块的12例;微钙化的10例;其他类型的10例)和68例正常对照。所有乳腺X线照片均使用计算机X线摄影(CR;采样间距为50μm)获得。12名乳腺X线摄影技师以随机顺序独立评估以硬拷贝和软拷贝形式呈现的CR乳腺X线照片,间隔至少4周。观察者在七点(1至7)和连续(0至100)恶性程度量表上对图像进行评分。进行了受试者操作特征分析,并计算了每种模式的平均曲线下面积(AUC)。将带有Bonferroni校正的留一法应用于多读者/多病例分析。在七点量表上,3-MP LCD、5-MP LCD和硬拷贝胶片的平均AUC值分别为0.954、0.947和0.956;在连续量表上分别为0.943、0.923和0.944。在两种量表上,三种模式之间均无显著差异。使用3-MP和5-MP LCD进行软拷贝阅读在检测乳腺癌方面与硬拷贝阅读相当。