Junsomboon Jaroon, Jakmunee Jaroon
Department of Chemistry, Faculty of Science, Chiang Mai University, Chiang Mai 50200, Thailand.
Talanta. 2008 Jul 15;76(2):365-8. doi: 10.1016/j.talanta.2008.03.006. Epub 2008 Mar 16.
A simple flow injection system using three 3-way solenoid valves as an electric control injection valve and with a simple home-made chloride ion selective electrode based on Ag/AgCl wire as a sensor for determination of water soluble chloride in admixtures and aggregates for cement has been developed. A liquid sample or an extract was injected into a water carrier stream which was then merged with 0.1M KNO(3) stream and flowed through a flow cell where the solution will be in contact with the sensor, producing a potential change recorded as a peak. A calibration graph in range of 10-100 mg L(-1) was obtained with a detection limit of 2 mg L(-1). Relative standard deviations for 7 replicates injecting of 20, 60 and 90 mg L(-1) chloride solutions were 1.0, 1.2 and 0.6%, respectively. Sample throughput of 60 h(-1) was achieved with the consumption of 1 mL each of electrolyte solution and water carrier. The developed method was validated by the British Standard methods.
已开发出一种简单的流动注射系统,该系统使用三个三通电磁阀作为电控注射阀,并采用基于Ag/AgCl丝的简易自制氯离子选择性电极作为传感器,用于测定水泥外加剂和集料中的水溶性氯化物。将液体样品或提取物注入水流载体流中,然后与0.1M KNO₃流合并,流经流通池,在流通池中溶液与传感器接触,产生的电位变化记录为一个峰。获得了10 - 100 mg L⁻¹范围内的校准曲线,检测限为2 mg L⁻¹。对20、60和90 mg L⁻¹氯化物溶液进行7次重复进样的相对标准偏差分别为1.0%、1.2%和0.6%。每次消耗1 mL电解质溶液和水流载体,实现了60 h⁻¹的样品通量。所开发的方法通过英国标准方法进行了验证。