Suppr超能文献

与临床方法和放射成像相比,在重症监护患者中使用比色二氧化碳传感器进行鼻肠饲管置入

Use of a colorimetric carbon dioxide sensor for nasoenteric feeding tube placement in critical care patients compared with clinical methods and radiography.

作者信息

Munera-Seeley Verónica, Ochoa Juan B, Brown Nefertiti, Bayless Angela, Correia M Isabel T D, Bryk Jodie, Zenati Mazen

机构信息

Department of Surgery, University of Pittsburgh Medical Center, Pittsburgh, PA 15213, USA.

出版信息

Nutr Clin Pract. 2008 Jun-Jul;23(3):318-21. doi: 10.1177/0884533608318105.

Abstract

BACKGROUND

Misplacement of nasoenteric feeding tubes (NFTs) into the airway instead of the esophagus leads to complications. Healthcare providers have relied on clinical methods, devices such as carbon dioxide (CO(2)) sensors, and radiography (the gold standard) to evaluate NFT placements. Most institutions include radiographs in their protocols for NFT insertions, making it expensive and cumbersome. A new commercial CO(2) sensor was developed to assist in these procedures, and the authors evaluated its use.

METHODS

Nurses performing NFT placement completed questionnaires following each procedure. The nurses recorded the clinical methods used to determine proper insertion and, based on them, where the NFT was located. Nurses then evaluated NFT insertion with the CO(2) sensor; from the readings, they recorded where the tube was located. Confirmation of tube placement was performed radiographically.

RESULTS

The authors evaluated 424 NFT insertions. Of these, 15 (3.5%) were incorrectly placed into the airway, and 409 were correctly placed into the esophagus. The CO(2) sensor correctly assessed NFT placement in 421 (99%) of the 424 cases. The authors found the device to have a sensitivity of 86.7% and a specificity of 99.8%.

CONCLUSIONS

The CO(2) sensor is a helpful bedside tool to use in conjunction with clinical methods during NFT insertions. However, there is insufficient evidence to abandon the use of radiographs to confirm tube placement.

摘要

背景

鼻肠饲管(NFT)误置入气道而非食管会导致并发症。医疗服务提供者一直依赖临床方法、诸如二氧化碳(CO₂)传感器之类的设备以及放射成像(金标准)来评估NFT的放置位置。大多数机构在其NFT插入操作规范中都包括放射成像,这使得操作既昂贵又繁琐。一种新型商用CO₂传感器被开发出来以辅助这些操作,作者对其使用情况进行了评估。

方法

进行NFT放置操作的护士在每次操作后填写问卷。护士记录用于确定正确插入位置的临床方法,并据此记录NFT的位置。然后护士使用CO₂传感器评估NFT的插入情况;根据读数记录导管的位置。通过放射成像确认导管放置情况。

结果

作者评估了424次NFT插入操作。其中,15次(3.5%)误置入气道,409次正确置入食管。CO₂传感器在424例中的421例(99%)中正确评估了NFT的放置位置。作者发现该设备的灵敏度为86.7%,特异性为99.8%。

结论

在NFT插入过程中,CO₂传感器是一种有助于与临床方法结合使用的床边工具。然而,尚无足够证据放弃使用放射成像来确认导管放置情况。

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验