Cheek Dennis, Gibbons John P, Rosen Isaac I, Hogstrom Kenneth R
Mary Bird Perkins Cancer Center, Baton Rouge, Louisiana 70809, USA.
Med Phys. 2008 Aug;35(8):3565-73. doi: 10.1118/1.2952362.
Helical tomotherapy is a technique for delivering intensity modulated radiation therapy treatments using a continuously rotating linac. In this approach, fan beams exiting the linac are dynamically modulated in synchrony with the motion of the gantry and couch. Helical IMRT deliveries have been applied to treating surface lesions, and the purpose of this study was to evaluate the accuracy of dose calculated by the TomoTherapy HiArt treatment planning system for superficial planning target volumes (PTVs). TomoTherapy treatment plans were developed for three superficial PTVs (2-, 4-, and 6-cm deep radially by 90 degrees azimuthally by 4-cm longitudinally) contoured on a 27-cm diameter cylindrical white opaque, high-impact polystyrene phantom. The phantom included removable transverse and sagittal film cassettes that contained bare Kodak EDR2 films cut such that their edges matched the phantom surface (+/-0.05 cm). The phantom was aligned to the machine's isocenter (+/-0.05 cm) and was irradiated according to the treatment plans. Films were scanned with a Vidar film digitizer, and optical densities were converted to dose using a calibration determined from a 6 MV perpendicular film exposure. This perpendicular calibration required that axial film doses (parallel irradiation) be scaled by 1.02 so that mid-arc depth doses matched those measured in the sagittal plane (perpendicular irradiation). All film readings were scaled by 0.935 to correct for over-response due to phantom Cerenkov light. Measured dose distributions were registered to calculated ones and compared. Calculated doses overpredicted measured doses by as much as 9.5% of the prescribed dose at depths less than 1 cm. At depths greater than 1 cm, calculated dose distributions showed agreement to measurement within 5% in the high-dose region and within 0.2 cm distance-to-agreement in the dose falloff regions. In the low-dose region posterior to the PTVs (<10% of the prescribed dose), the dose algorithm underpredicted the dose by 1%-2% of the prescribed dose. Clinically, it is recommended that 1 cm of bolus be used on the surface to ensure that cancerous tissues less than 1 cm depth are not underdosed.
螺旋断层放射治疗是一种利用连续旋转直线加速器进行调强放射治疗的技术。在这种方法中,离开直线加速器的扇形束与机架和治疗床的运动同步进行动态调制。螺旋调强放射治疗已应用于治疗体表病变,本研究的目的是评估TomoTherapy HiArt治疗计划系统针对浅表计划靶体积(PTV)计算剂量的准确性。针对在直径27厘米的圆柱形白色不透明、高抗冲聚苯乙烯体模上勾勒出的三个浅表PTV(径向深度分别为2厘米、4厘米和6厘米,方位角90度,纵向4厘米)制定了TomoTherapy治疗计划。该体模包括可移动的横向和矢状胶片暗盒,其中装有裁切后的柯达EDR2裸片,使其边缘与体模表面匹配(±0.05厘米)。将体模对准机器的等中心(±0.05厘米),并根据治疗计划进行照射。用Vidar胶片数字化仪扫描胶片,并使用由6兆伏垂直胶片曝光确定的校准将光学密度转换为剂量。这种垂直校准要求轴向胶片剂量(平行照射)按1.0进行缩放,以使弧中心深度剂量与矢状面(垂直照射)中测量的剂量相匹配。所有胶片读数均按0.935进行缩放,以校正由于体模切伦科夫光导致的过度响应。将测量的剂量分布与计算的剂量分布进行配准并比较。在深度小于1厘米时,计算剂量比测量剂量高估多达规定剂量的9.5%。在深度大于1厘米时,计算的剂量分布在高剂量区域与测量值的一致性在5%以内,在剂量下降区域距离一致性在0.2厘米以内。在PTV后方的低剂量区域(<规定剂量的10%),剂量算法预测的剂量比规定剂量低1%-2%。临床上,建议在表面使用1厘米厚的等效物,以确保深度小于1厘米的癌组织不会剂量不足。