Hoburg Arnd T, Mehta Manav, Tohtz Stephan, Perka Carsten
Department of Orthopaedics, Trauma and Reconstructive Surgery, Humboldt University of Berlin, Schumannstrasse 20/21, Charité, 10117, Berlin, Germany.
Clin Orthop Relat Res. 2009 Mar;467(3):850-4. doi: 10.1007/s11999-008-0526-7. Epub 2008 Sep 27.
The use of cementless threaded cups in THA is a well-established treatment. Fractures of the cups are rare complications recorded in individual cases with material defects being discussed as the primary cause. We analyzed three cases of fractured cups. Although all three cups were well fixed to existing bone, we observed deficient osseous backing dorsocranially and abrasion particles. There were no signs of femoroacetabular impingement or infection. The cups showed corrosion debris. Scanning electron microscopic investigations showed characteristics of fretting and fretting-related corrosion. We concluded the fractures occurred because of fretting combined with inadequate bony support leading to fatigue of the material and subsequent fracture.
在全髋关节置换术中使用非骨水泥螺纹髋臼杯是一种成熟的治疗方法。髋臼杯骨折是个别病例中记录的罕见并发症,材料缺陷被认为是主要原因。我们分析了三例髋臼杯骨折病例。尽管所有三个髋臼杯都与现存骨牢固固定,但我们观察到背颅侧骨质支撑不足和磨损颗粒。没有股骨髋臼撞击或感染的迹象。髋臼杯显示出腐蚀碎片。扫描电子显微镜检查显示有微动和微动相关腐蚀的特征。我们得出结论,骨折是由于微动加上骨质支撑不足导致材料疲劳并随后骨折。