Ciszewski A, Lukaszewski Z
Institute of General Chemistry, Technical University of Poznań, 60-965 Poznań, Poland.
Talanta. 1983 Nov;30(11):873-5. doi: 10.1016/0039-9140(83)80202-1.
The determination of trace levels of thallium in lead and lead salts by differential pulse anodic-stripping voltammetry has been made possible by using a surfactant as an electrochemical masking agent in addition to a complexing agent. In 0.2M EDTA at pH 4.5 as supporting electrolyte without surfactant, lead at concentrations below 0.5mM does not give a peak. When the electrolyte also contains tetrabutylammonium chloride (TBAC) at 0.01 M concentration, lead can be tolerated at concentrations up to 0.05M, while the height of the thallium peak is unaffected. It is thus possible to determine 5nM T1(I) in the presence of 0.05M Pb(II), i.e., Tl at the 1 x 10(-5)% level in lead. The precision of the determination (1-4%) and the recovery are satisfactory. Neither an 800-fold excess ratio of Cu(II) to Tl(I) nor a 10(7)-fold ratio of Bi(III) interferes in the determination. Thallium has been determined in a range of lead salts of various degrees of purity.