Knochen Moisés, Giglio Javier
Universidad de la República, Facultad de Quimica, Departamento "Estrella Campos", Av. Gral. Flores 2124. Casilla de Correo 1157, 11800 Montevideo, Uruguay.
Talanta. 2004 Dec 15;64(5):1226-32. doi: 10.1016/j.talanta.2004.03.071.
A flow injection method is proposed for the determination of phenylephrine hydrochloride in pharmaceutical dosage forms. The method involves the use of on-line solid-phase extraction by means of a microcolumn containing Dowex 50W X8 ion-exchange resin for the separation of the analyte prior to colour development and spectrophotometric detection in the visible region. The influence of preconcentration flow, preconcentration pH and elution volume was studied. The method exhibits appropriate linearity (r(2) = 0.9999) which was proved statistically by means of the "F"-test. When applied to commercial samples containing several active ingredients and excipients, a significant reduction of interferences was found. Accuracy, evaluated by means of the spike recovery method was in the range 99.7-100.8%, with precision (R.S.D., %) better than 1%. In order to achieve the automation the system was controlled from a notebook computer by means of a program written in QuickBASIC language. Under these conditions, a sampling frequency of 40 samples per hour could be attained.
本文提出了一种流动注射法用于测定药物剂型中的盐酸去氧肾上腺素。该方法包括在显色和可见区分光光度检测之前,通过含有Dowex 50W X8离子交换树脂的微柱进行在线固相萃取来分离分析物。研究了预富集流速、预富集pH值和洗脱体积的影响。该方法具有适当的线性(r(2) = 0.9999),通过“F”检验在统计学上得到了证实。当应用于含有多种活性成分和辅料的商业样品时,发现干扰显著减少。通过加标回收法评估的准确度在99.7 - 100.8%范围内,精密度(相对标准偏差,%)优于1%。为了实现自动化,该系统通过用QuickBASIC语言编写的程序由笔记本电脑控制。在这些条件下,每小时可达到40个样品的采样频率。