Rivaz Hassan, Fleming Ioana, Assumpcao Lia, Fichtinger Gabor, Hamper Ulrike, Choti Michael, Hager Gregory, Boctor Emad
ERC for Computer Integrated Surgery, Johns Hopkins University, USA.
Med Image Comput Comput Assist Interv. 2008;11(Pt 2):458-66. doi: 10.1007/978-3-540-85990-1_55.
The clinical feasibility of 2D elastography methods is hindered by the requirement that the operator avoid out-of-plane motion of the ultrasound image during palpation, and also by the lack of volumetric elastography measurements. In this paper, we develop and evaluate a 3D elastography method operating on volumetric data acquired from a 3D probe. Our method is based on minimizing a cost function using dynamic programming (DP). The cost function incorporates similarity of echo amplitudes and displacement continuity. We present, to the best of our knowledge, the first in-vivo patient studies of monitoring liver ablation with freehand DP elastography. The thermal lesion was not discernable in the B-mode image but it was clearly visible in the strain image as well as in validation CT. We also present 3D strain images from thermal lesions in ex-vivo ablation. Good agreement was observed between strain images, CT and gross pathology.