轻量可乐、磷酸和柠檬酸对弹性体链的环境影响。

The environmental influence of Light Coke, phosphoric acid, and citric acid on elastomeric chains.

作者信息

Teixeira Leandro, Pereira Betina do Rosario, Bortoly Thais Gelatti, Brancher Joao Armando, Tanaka Orlando Motohiro, Guariza-Filho Odilon

机构信息

Orthodontics, Pontifical Catholic University of Paraná in Curitiba, Brazil.

出版信息

J Contemp Dent Pract. 2008 Nov 1;9(7):17-24.

Abstract

AIM

The purpose of this study was to evaluate the in vitro effect of Light Coke, phosphoric acid, and citric acid on the force decline pattern of two types of elastomeric chains.

METHODS AND MATERIALS

One hundred sixty gray colored elastomeric chain modules, 80 Chainette (GAC) and 80 Sunburst (GAC) chains, were divided into four groups for immersion into Light Coke, phosphoric acid, citric acid, and artificial saliva. The initial stretched force of the elastomeric chains ranged from 220 gf to 250 gf. During the experiment, all elastomeric chains were kept immersed in artificial saliva at 37 masculineC (pH 6.24) to simulate the oral environment. To simulate daily consumption of Light Coke, the elastomeric chains were immersed in the solutions twice a day for 15 minutes. The control group was kept immersed in artificial saliva continuously with no further treatment. Force (gf) was measured with a Dial-Type dynamometer (Dentaurum), initially (baseline), 24 hours, 7, 14, and 21 days. Data were analyzed using the Tukey test at a 5% level of probability.

RESULTS

A statistically significant reduction on the force produced by the elastomeric chains was seen at different time points. The greatest reduction in force occurred in the first 24 hours (p=0.01).

CONCLUSION

Force decay of the Chainette elastomeric chains were lower than the Sunburst (p<0.05). The immersion treatments caused no statistically significant difference in force for either chain module (p>0.05).

CLINICAL SIGNIFICANCE

Good elastomeric chain properties are necessary for effective tooth movement, and knowledge of force-decay rates is an important factor in achieving the best orthodontic outcome.

摘要

目的

本研究旨在评估健怡可乐、磷酸和柠檬酸对两种类型弹性链的力衰减模式的体外影响。

方法和材料

160个灰色弹性链模块,80个链状(GAC)和80个旭日状(GAC)链,被分为四组,分别浸泡在健怡可乐、磷酸、柠檬酸和人工唾液中。弹性链的初始拉伸力范围为220克力至250克力。在实验过程中,所有弹性链都保持浸泡在37摄氏度(pH 6.24)的人工唾液中以模拟口腔环境。为模拟日常饮用健怡可乐的情况,弹性链每天浸泡在溶液中两次,每次15分钟。对照组持续浸泡在人工唾液中,不进行进一步处理。使用表盘式测力计(Dentaurum)在初始(基线)、24小时、7天、14天和21天测量力(克力)。数据使用Tukey检验在5%的概率水平上进行分析。

结果

在不同时间点观察到弹性链产生的力有统计学上的显著降低。力的最大降低发生在最初的24小时内(p = 0.01)。

结论

链状弹性链的力衰减低于旭日状弹性链(p < 0.05)。浸泡处理对两种链模块的力没有统计学上的显著差异(p > 0.05)。

临床意义

良好的弹性链性能对于有效的牙齿移动是必要的,而了解力衰减率是实现最佳正畸效果的一个重要因素。