Tsang Susanna, Phu Frank, Baum Marc M, Poskrebyshev Gregory A
Oak Crest Institute of Science, 2275 E. Foothill Blvd., Pasadena, CA 91107, United States.
Talanta. 2007 Mar 15;71(4):1560-8. doi: 10.1016/j.talanta.2006.07.043. Epub 2006 Aug 28.
A modified molybdenum blue method for fast and accurate measurement of arsenate and phosphate in aqueous solution at concentrations below 10mumoll(-1) is reported. The modification consists of optimizing the composition of the molybdenum-containing solution (potassium antimony tartrate, ammonium molybdate, sulfuric acid). Selective reduction of arsenate by sodium dithionite is used to determine phosphate concentrations, and for the speciation of arsenate and arsenite, in an aqueous mixture according to the scheme developed by Johnston and Pilson. Sodium dithionite is used for the first time to achieve complete, fast (<10min), and selective reduction of arsenate in neutral solution. These two significant improvements afforded a colorimetric limit for As detection near 1ppb, which easily meets the requirements imposed by the revised EPA threshold levels for As in drinking water.
报道了一种改进的钼蓝法,用于快速准确地测量浓度低于10μmol L⁻¹的水溶液中的砷酸盐和磷酸盐。改进之处在于优化含钼溶液(酒石酸锑钾、钼酸铵、硫酸)的组成。根据约翰斯顿和皮尔森提出的方案,用连二亚硫酸钠选择性还原砷酸盐来测定磷酸盐浓度,并对水混合物中的砷酸盐和亚砷酸盐进行形态分析。连二亚硫酸钠首次用于在中性溶液中实现砷酸盐的完全、快速(<10分钟)和选择性还原。这两项重大改进使砷检测的比色法检测限接近1 ppb,轻松满足了美国环境保护局修订后的饮用水中砷阈值水平的要求。