Electron beam effective source surface distances for a high energy linear accelerator.

作者信息

Sharma S C, Johnson M W

机构信息

Department of Radiation Oncology, E. W. Sparrow Hospital, Lansing, MI 48909.

出版信息

Med Dosim. 1991 Jun;16(2):65-70. doi: 10.1016/0958-3947(91)90048-7.

Abstract

The design of the Varian Clinac 1800 linear accelerator electron applicator system does not allow clearance for all head and neck patients to be treated at the standard calibration distance of 100 cm. Discrepancies have been found between dose calculations using the inverse square law for extended distances and their measured data. A 4 X 4 cm2 applicator at an energy of 9 MeV, for example, had dose differences of 13 and 23% at distances of 105 and 110 cm SSD. Because of these discrepancies, effective source surface distances (SSDeff) were determined for all the standard electron energies and applicators of a Clinac 1800. These effective source surface distances ranged from 41.6 cm to 92.6 cm for the 4 X 4 cm2 cone/6 MeV electron beam through the 25 X 25 cm2 cone/20 MeV electron beam. A summary of these distances and an analysis of the clinical use of both a best fit SSDeff and a common SSDeff for patient dosimetry calculations is presented.