Shi Changyan, Xie Shaoai, Jia Jinping
School of Environmental Science and Engineering, Shanghai Jiao Tong University, Shanghai 200240, China.
J Autom Methods Manag Chem. 2008;2008:453429. doi: 10.1155/2008/453429. Epub 2008 Dec 14.
A new method of indirect determination of Cu(2+) was developed based on square-wave voltammetry by the oxidation of iodide in organic solvent at the liquid/liquid (L/L) interface. The limit of detection for the determination of Cu(2+) in this method was found to be 5 x 10(-4) mol/L, and the concentration ranged up to 1 x 10(-2) mol/L gave a linear limiting current versus concentration response. For the same simulated wastewater, this method showed high accuracy compared with the result tested by sodium diethyldithiocarbamate extraction spectrophotometry. This approach could be applied to the indirect determination of the oxidative heavy metals in the industrial wastewater.
基于方波伏安法,通过在液/液(L/L)界面的有机溶剂中氧化碘化物,开发了一种间接测定Cu(2+)的新方法。该方法测定Cu(2+)的检测限为5×10(-4) mol/L,浓度范围高达1×10(-2) mol/L时,极限电流与浓度呈线性响应。对于相同的模拟废水,与二乙基二硫代氨基甲酸钠萃取分光光度法测试结果相比,该方法具有较高的准确度。该方法可应用于工业废水中氧化性重金属的间接测定。