Olivé-Monllau Rosa, Orozco Jahir, Fernández-Sánchez César, Baeza Mireia, Bartrolí Jordi, Jimenez-Jorquera Cecilia, Céspedes Francisco
Grup de Sensors i Biosensors, Departament de Química, Facultat de Ciències, Edifici C-Nord, Universitat Autònoma de Barcelona, 08193 Bellaterra (Cerdanyola del Vallès), Spain.
Talanta. 2009 Mar 15;77(5):1739-44. doi: 10.1016/j.talanta.2008.10.015. Epub 2008 Oct 18.
This work reports on the performance of a user-friendly flow injection analysis (FIA) system for the monitoring of free chlorine. A methacrylate flow cell integrating a gold thin-film microelectrode, together with an on-chip gold counter electrode, both fabricated by microfabrication technology, provided robustness, low output impedance, rapid response and low cost to the proposed flow system. An external Ag/AgCl reference electrode placed downstream the chip completes the electrochemical cell. Amperometric detection of chlorine was carried out at a set potential of +350 mV, without oxygen interference. The proposed flow system responded linearly to chlorine concentrations in a range from 0.2 to 5 mgl(-1), with a sensitivity of 0.23 microAlmg(-1), the estimated limit of detection being 0.02 mgl(-1). In addition, the system response was kept stable for at least 10 days (+/-3sigma criterion), by keeping the flow system in an inert atmosphere when not in use. Fifteen samples of swimming pool waters were analyzed and no matrix effects were detected. Also, results were in good agreement with those obtained by a standard method. The excellent analytical performance of the system together with its good working stability would also enable its application for the detection of chlorine in other matrices such as tap water or chlorine stock solutions.
这项工作报道了一种用于监测游离氯的用户友好型流动注射分析(FIA)系统的性能。一个集成了金薄膜微电极的甲基丙烯酸酯流通池,以及一个同样通过微加工技术制造的片上金对电极,为所提出的流动系统提供了稳健性、低输出阻抗、快速响应和低成本。置于芯片下游的外部Ag/AgCl参比电极构成了完整的电化学池。在+350 mV的设定电位下进行氯的安培检测,不受氧气干扰。所提出的流动系统对氯浓度在0.2至5 mg l⁻¹范围内呈线性响应,灵敏度为0.23 μA mg⁻¹,估计检测限为0.02 mg l⁻¹。此外,当不使用时,通过将流动系统置于惰性气氛中,系统响应至少可保持10天稳定(±3σ标准)。分析了15个游泳池水样,未检测到基质效应。而且,结果与通过标准方法获得的结果高度一致。该系统出色的分析性能及其良好的工作稳定性也将使其能够应用于检测其他基质中的氯,如自来水或氯储备溶液。