Choi Hyun-Seok, Myoung Hyoun-Seok, Lee Hyo-Ki, Park Ho-Dong, Lee Kyoung-Joung
Department of Biomedical Engineering, Yonsei University, Republic of Korea.
Annu Int Conf IEEE Eng Med Biol Soc. 2008;2008:270-2. doi: 10.1109/IEMBS.2008.4649142.
In this paper, a new noise reduction method was proposed for oscillometric blood pressure measurement. The proposed method uses a capacitive sensor and an adaptive filter to minimize blood pressure measurement error. Noise such as undesired external pressure applied to cuff was focused on. Three types of the distorted oscillation signals (no overlap, non-consecutive overlap, consecutive overlap between the noise and the oscillation) were used to compare the conventional method using linear interpolation and the proposed method using the adaptive filter. The proposed method outperformed the conventional method in the case of consecutive overlap between the noise and the oscillation. The proposed method could be useful for measuring blood pressure in such a noisy environment that the subject is being transported.
本文提出了一种用于示波法血压测量的新型降噪方法。该方法采用电容式传感器和自适应滤波器来最小化血压测量误差。重点关注施加在袖带上的不期望的外部压力等噪声。使用三种类型的失真振荡信号(无重叠、非连续重叠、噪声与振荡之间的连续重叠)来比较使用线性插值的传统方法和使用自适应滤波器的所提出的方法。在所提出的方法在噪声与振荡之间存在连续重叠的情况下优于传统方法。所提出的方法对于在诸如患者正在被转运的嘈杂环境中测量血压可能是有用的。