Taralunga Dragos D, Wolf Werner, Strungaru Rodica, Ungureanu G
Politehnica University of Bucharest, Electronics, Telecommunications, and Information Technology Faculty, Iuliu Maniu 1-3, RO-061072 Bucharest, Romania.
Annu Int Conf IEEE Eng Med Biol Soc. 2008;2008:5402-5. doi: 10.1109/IEMBS.2008.4650436.
Abdominal signals (ADS) recorded from pregnant women represent an important tool for monitoring the fetal heart rate (FHR) variability and the well-being state of the fetus, mainly because it has the advantage of being noninvasive. Thus, no risk is given during recording either for the mother or for the fetus, but complex signal processing steps are necessary, mainly due to the presence of the maternal ECG in the ADS, in order to achieve a clean fetal electrocardiogram (fECG). The paper presents an improved application of the Event Synchronous Canceller (ESC) for maternal electrocardiogram (mECG) suppression. An adaptive mECG template which also includes the P and T waves is considered for the ESC algorithm. ESC is able to perfectly separate the mECG even though the fetal beats overlap with the maternal QRS complex (mQRS). The algorithm is applied both on real ADS recorded during pregnancy and on simulated ADS data; the latter now uses simulation of all signal and noise components, including the fetal and maternal ECG. The modified ESC shows good results in extracting a cleaned fECG signal that can help in reducing the inconsistency in interpretation of FHR. Thus the false positive diagnosis regarding the health status of fetus is strongly reduced.
从孕妇身上记录的腹部信号(ADS)是监测胎儿心率(FHR)变异性和胎儿健康状况的重要工具,主要是因为它具有非侵入性的优势。因此,在记录过程中对母亲和胎儿都没有风险,但由于ADS中存在母体心电图,需要进行复杂的信号处理步骤,以获得清晰的胎儿心电图(fECG)。本文提出了一种改进的事件同步消除器(ESC)用于抑制母体心电图(mECG)的应用。ESC算法考虑了一个自适应的mECG模板,该模板还包括P波和T波。即使胎儿心跳与母体QRS复合波(mQRS)重叠,ESC也能完美地分离mECG。该算法应用于孕期记录的真实ADS和模拟ADS数据;后者现在使用所有信号和噪声成分的模拟,包括胎儿和母体心电图。改进后的ESC在提取清洁的fECG信号方面显示出良好的效果,有助于减少FHR解释中的不一致性。因此,关于胎儿健康状况的假阳性诊断大大减少。