Herpers Bjorn L, de Jong Ben A W, Dekker Bertie, Aerts Piet C, van Dijk Hans, Rijkers Ger T, van Velzen-Blad Heleen
Department of Medical Microbiology and Immunology, St. Antonius Hospital, Nieuwegein, The Netherlands.
J Immunol Methods. 2009 Mar 31;343(1):61-3. doi: 10.1016/j.jim.2009.01.002. Epub 2009 Jan 24.
Diagnostic assays for measurement of functional mannose-binding lectin (MBL) in serum are widely performed as part of immune status assessment. Classical pathway mediated complement activity can interfere in these functional MBL assays. Here we describe classical pathway interference incidentally occurring in a previously described hemolytic MBL assay and the modification of this assay to prevent this artifact by addition of anti-C1q antibodies. Classical pathway interference in functional MBL assays can and should be inhibited to prevent that MBL deficiency is overlooked and patients are misdiagnosed.
作为免疫状态评估的一部分,血清中功能性甘露糖结合凝集素(MBL)的诊断检测被广泛开展。经典途径介导的补体活性会干扰这些功能性MBL检测。在此,我们描述了在先前描述的溶血MBL检测中偶然出现的经典途径干扰,以及通过添加抗C1q抗体对该检测进行的改进,以防止出现这种假象。功能性MBL检测中的经典途径干扰能够且应该被抑制,以防止MBL缺乏被忽视以及患者被误诊。