J Pharm Biomed Anal. 2009 Apr 5;49(3):594-600. doi: 10.1016/j.jpba.2008.10.038. Epub 2008 Nov 5.

Abstract

The paper describes the development and validation of a new derivatization-free liquid chromatography method for simultaneous determination of propylene glycol and formaldehyde in the formulations containing formaldehyde-releasing preservative. Highly swollen hydrogel made of poly(ethylene glycol)-protein conjugates was taken as a model formulation for integration of the propylene glycol and the diazolydinyl urea as formaldehyde releaser. The method is shown to be simple and selective and, more importantly, allows determining an existing level of formaldehyde at the moment of analysis instead of all available formaldehyde that might be released during chemical derivatization. After liquid extraction the propylene glycol (PG) and formaldehyde (FA) amounts are determined chromatographically on a Shodex SH 1011 ligand-exchange column using 0.01 M sulfuric acid mobile phase, a flow rate of 1.0 ml/min and RI detection. The assay is validated showing good linearity, precision, and accuracy. The limits of detection of formaldehyde and propylene glycol in the analyzed solutions were estimated to be 25 ng and 87 ng, respectively. This analytical assay is considered useful for product stability studies and in developing new formaldehyde releaser-containing formulations where the concentration of formaldehyde is a presumable subject of labeling requirements. This method can also provide a rapid and convenient alternative to gas chromatography method of propylene glycol quantification.

摘要

本文描述了一种用于同时测定含甲醛释放防腐剂制剂中丙二醇和甲醛的新型无衍生化液相色谱方法的开发与验证。由聚(乙二醇)-蛋白质缀合物制成的高度溶胀水凝胶被用作丙二醇与作为甲醛释放剂的重氮烷基脲整合的模型制剂。该方法简单且具有选择性,更重要的是,它能够在分析时测定甲醛的现有水平,而不是化学衍生化过程中可能释放的所有可用甲醛。液体萃取后,丙二醇(PG)和甲醛(FA)的含量在Shodex SH 1011配体交换柱上进行色谱测定,使用0.01 M硫酸作为流动相,流速为1.0 ml/min,采用示差折光检测。该测定方法经过验证,显示出良好的线性、精密度和准确度。分析溶液中甲醛和丙二醇的检测限分别估计为25 ng和87 ng。这种分析方法被认为对产品稳定性研究以及开发含甲醛释放剂的新制剂有用,在这些制剂中甲醛浓度可能是标签要求的主题。该方法还可以为丙二醇定量的气相色谱法提供一种快速便捷的替代方法。

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验