Hoizey Guillaume, Lamiable Denis, Gozalo Claire, Miric Thomas, Thomas Aurélie, Binet Laurent, Oget Olivier, Millart Hervé, Mauprivez Cédric
Laboratoire de Pharmacologie et Toxicologie, CHU de Reims, Reims, France.
J Pharm Biomed Anal. 2009 May 1;49(4):1082-7. doi: 10.1016/j.jpba.2009.01.028. Epub 2009 Jan 31.
A specific liquid chromatography-mass spectrometric (LC-MS) method using an ion trap spectrometer was developed for the quantitation of articaine in human plasma. Articaine and the internal standard (trazodone) were extracted in a single step with diethyl-ether from 0.5 mL of alkalinized plasma. The mobile phase consisted of acetonitrile with 0.1% formic acid (40:60, v/v). It was delivered at a flow rate of 0.3 mL/min. The effluent was monitored by MS in positive-ion mode. Ionisation was performed using an electrospray ion source operating at 200 degrees C. Articaine was identified and quantified in SIM mode at m/z 185. Calibration curves were linear over the concentration range of 78.1-5000 ng/mL with determination coefficients>0.996. This method was fast (total run-time<3 min), accurate (bias<16%), and reproducible (intra-assay and inter-assay precision<14%) with a quantitation limit of 78.1 ng/mL. The good specificity and sensitivity achieved by this method allowed the determination of articaine plasma levels in patients following a submucosal infiltration injection of articaine in the patients undergoing a third molar surgery.