Scoliosis in-brace curve correction and patient preference of CAD/CAM versus plaster molded TLSOs.

作者信息

Sankar Wudbhav N, Albrektson Josh, Lerman Lawrence, Tolo Vernon T, Skaggs David L

机构信息

Children's Hospital Los Angeles, 4650 Sunset Boulevard #69, Los Angeles, CA, 90027, USA.

出版信息

J Child Orthop. 2007 Dec;1(6):345-9. doi: 10.1007/s11832-007-0066-9. Epub 2007 Nov 21.

Abstract

PURPOSE

CAD/CAM technology is a newer technique for creating spinal orthoses than standard plaster molded methods. To our knowledge there has been only one previous study of CAD/CAM braces. The purpose of our study was to compare patient preference and in-brace correction of Cobb angle between plaster molded thoracolumbosacral orthoses (TLSO) and CAD/CAM designed TLSOs in a series of patients with scoliosis.

METHODS

Ten patients with an average initial Cobb angle of 30.8 degrees (range 18 degrees -46 degrees ) had both a plaster molded TLSO and a CAD/CAM TLSO fabricated for them. In each case, the decision to brace was made by the treating surgeon based on curve magnitude and skeletal maturity. After 3 weeks of 23 h a day wear, in-brace correction of the Cobb angle was measured for each brace based on standard PA spine radiographs. After 3 months of use, patients were asked which brace they preferred.

RESULTS

For the CAD/CAM brace, the mean curve correction after 3 months was 51% compared to 44% in the plaster molded TLSO cohort. (p = 0.46). Seven out of nine patients preferred the CAD/CAM TLSO over the plaster molded TLSO. There were no brace complications in either group.

CONCLUSION

In our matched cohort study, CAD/CAM TLSOs had at least equivalent if not superior correction of the Cobb angle compared to standard plaster molded TLSOs; 78% of our patients preferred the CAD/CAM brace over the standard TLSO.

摘要

目的

与标准石膏塑形方法相比,计算机辅助设计/计算机辅助制造(CAD/CAM)技术是一种用于制作脊柱矫形器的新技术。据我们所知,此前仅有一项关于CAD/CAM支具的研究。我们研究的目的是比较一系列脊柱侧弯患者中,石膏塑形胸腰骶部矫形器(TLSO)与CAD/CAM设计的TLSO在患者偏好以及支具内Cobb角矫正方面的情况。

方法

10例平均初始Cobb角为30.8度(范围18度 - 46度)的患者分别制作了石膏塑形TLSO和CAD/CAM TLSO。在每种情况下,治疗外科医生根据侧弯程度和骨骼成熟度决定是否使用支具。每天佩戴23小时,持续3周后,根据标准脊柱正位X线片测量每个支具的Cobb角在支具内的矫正情况。使用3个月后,询问患者更喜欢哪种支具。

结果

对于CAD/CAM支具,3个月后的平均侧弯矫正率为51%,而石膏塑形TLSO组为44%。(p = 0.46)。9例患者中有7例更喜欢CAD/CAM TLSO而非石膏塑形TLSO。两组均未出现支具相关并发症。

结论

在我们的配对队列研究中,与标准石膏塑形TLSO相比,CAD/CAM TLSO在Cobb角矫正方面即使不更优也至少相当;78%的患者更喜欢CAD/CAM支具而非标准TLSO。

相似文献

引用本文的文献

2
Scoliosis Bracing: Details Make the Difference.脊柱侧弯支具治疗:细节决定成败。
J Pediatr Soc North Am. 2024 Feb 5;5(4):776. doi: 10.55275/JPOSNA-2023-776. eCollection 2023 Nov.
8

本文引用的文献

文献检索

告别复杂PubMed语法,用中文像聊天一样搜索,搜遍4000万医学文献。AI智能推荐,让科研检索更轻松。

立即免费搜索

文件翻译

保留排版,准确专业,支持PDF/Word/PPT等文件格式,支持 12+语言互译。

免费翻译文档

深度研究

AI帮你快速写综述,25分钟生成高质量综述,智能提取关键信息,辅助科研写作。

立即免费体验