激光荧光检测恒牙磨牙咬合面牙本质龋损的性能:一项对样本进行全面验证的体内研究。

Performance of laser fluorescence for detection of occlusal dentinal caries lesions in permanent molars: an in vivo study with total validation of the sample.

作者信息

Abalos C, Herrera M, Jiménez-Planas A, Llamas R

机构信息

Department of Dental Materials, University of Seville, Seville, Spain.

出版信息

Caries Res. 2009;43(2):137-41. doi: 10.1159/000209347. Epub 2009 Mar 24.

Abstract

The objective of this in vivo study was to evaluate the performance of a laser fluorescence device in detecting dentinal caries in noncavitated occlusal surfaces. The sample included 102 first and second permanent molars, selected according to the criteria of Ekstrand et al. [Caries Res 1997;31:224-231]: 46 not suspected of having dentinal caries (score: 0-2) and 56 under suspicion (score: 3-4). Once measured by laser fluorescence, all were validated by fissurotomy (gold standard). To avoid ethical questions, the 46 teeth (score: 0-2) used were to serve as abutments for a fixed-fixed bridge. Visual inspection performed well, showing a sensitivity and specificity of 0.93 and 0.88, respectively. Laser fluorescence had an area under the receiver operating characteristic curve of A(z) = 0.85. The cutoff point with the highest sensitivity and specificity was 20, with a sensitivity and specificity of 0.89 and 0.75, respectively.

摘要

这项体内研究的目的是评估一种激光荧光装置在检测非龋洞性咬合面牙本质龋方面的性能。样本包括102颗第一恒磨牙和第二恒磨牙,根据埃克斯特兰德等人的标准[《龋病研究》1997年;31:224 - 231]选取:46颗未怀疑有牙本质龋(评分:0 - 2),56颗疑似有牙本质龋(评分:3 - 4)。经激光荧光测量后,所有牙齿均通过窝沟切开术(金标准)进行验证。为避免伦理问题,所使用的46颗牙齿(评分:0 - 2)用作固定桥的桥基牙。视觉检查表现良好,灵敏度和特异度分别为0.93和0.88。激光荧光的受试者工作特征曲线下面积A(z)=0.85。灵敏度和特异度最高的截断点为20,灵敏度和特异度分别为0.89和0.75。

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验