Bone quality assessment based on cone beam computed tomography imaging.

作者信息

Hua Yan, Nackaerts Olivia, Duyck Joke, Maes Frederik, Jacobs Reinhilde

机构信息

Oral Imaging Center, School of Dentistry, Oral Pathology and Maxillofacial Surgery, Faculty of Medicine, Katholieke Universiteit Leuven, Leuven, Belgium.

出版信息

Clin Oral Implants Res. 2009 Aug;20(8):767-71. doi: 10.1111/j.1600-0501.2008.01677.x. Epub 2009 May 27.

Abstract

OBJECTIVES

The aim of this in vitro study was to investigate the accuracy of fractal analysis and morphometry for bone quality assessment as measured with dual energy X-ray absorptiometry (DXA).

MATERIAL AND METHODS

Nineteen mandibular bone samples were used for the creation of artificial bone lesions (n=5) or decalcification (n=12) to simulate osteoporosis; two samples were used as controls. Cone beam computed tomography (CBCT) and DXA scans were made before and after processing the samples. The image data obtained from the CBCT scans were used to calculate the mean fractal dimension (FD), bone area and density (morphometric analysis) of the samples. Bone mineral density (BMD) was obtained from the DXA scans and set as a reference value for bone quality. The correlation between BMD and FD and between BMD and morphometric results were calculated.

RESULTS

A significant correlation between FD and BMD (rho=+0.71 to +0.75; P<0.05) was observed. Bone area and BMD of the specimens (rho=+0.69 to +0.85; P<0.05) were also significantly related, in contrast to the density analysis, for which no significant correlation to BMD was found.

CONCLUSIONS

The results of this study suggest that fractal analysis and bone area measurement have potential to evaluate bone quality on CBCT images, while density measurement does not seem to be valid.

摘要

目的

本体外研究的目的是调查分形分析和形态测量法在通过双能X线吸收测定法(DXA)测量骨质量评估中的准确性。

材料与方法

使用19个下颌骨样本制造人工骨病变(n = 5)或脱钙(n = 12)以模拟骨质疏松症;两个样本用作对照。在处理样本之前和之后进行锥束计算机断层扫描(CBCT)和DXA扫描。从CBCT扫描获得的图像数据用于计算样本的平均分形维数(FD)、骨面积和密度(形态测量分析)。从DXA扫描获得骨矿物质密度(BMD)并将其设定为骨质量的参考值。计算BMD与FD之间以及BMD与形态测量结果之间的相关性。

结果

观察到FD与BMD之间存在显著相关性(rho = +0.71至+0.75;P<0.05)。样本的骨面积与BMD(rho = +0.69至+0.85;P<0.05)也显著相关,与之形成对比的是,密度分析与BMD未发现显著相关性。

结论

本研究结果表明,分形分析和骨面积测量在CBCT图像上具有评估骨质量的潜力,而密度测量似乎无效。