Bhardwaj Atul, Hollenbeak Christopher S, Pooran Nakechand, Mathew Abraham
Division of General Internal Medicine, Penn State Milton S. Hershey Medical Center, 500 University Drive, Hershey, PA 17033, USA.
Am J Gastroenterol. 2009 Jun;104(6):1533-9. doi: 10.1038/ajg.2009.86. Epub 2009 Apr 21.
To evaluate the diagnostic accuracy of esophageal capsule endoscopy (ECE) for Barrett's esophagus (BE) in patients with gastroesophageal reflux disease (GERD).
Literature was searched for blinded studies evaluating the diagnostic accuracy of ECE for BE in patients with GERD. Meta-analysis was carried out to calculate pooled sensitivity and specificity of ECE for diagnosis of BE. Subgroup analysis was also carried out based on the reference standard used.
Nine studies comprising a total of 618 patients met the inclusion criteria. The pooled sensitivity and specificity of ECE for the diagnosis of BE for all studies were 77 and 86% respectively. The pooled sensitivity and specificity of ECE for the diagnosis of BE using esophagogastroduodenoscopy (EGD) as the reference standard were 78 and 90%, respectively; using histologically confirmed intestinal metaplasia (IM) as the reference standard pooled sensitivity and specificity were 78 and 73%, respectively. Statistical heterogeneity was not evident among studies for sensitivity results (P=0.270, I(2)=19), but heterogeneity was present for specificity results (P<0.001, I(2)=74). There was no evidence of publication bias. The ECE was found to be safe and had a high rate of patient preference.
Capsule endoscopy of esophagus has a moderate sensitivity and specificity for the diagnosis of BE in patients with GERD. The EGD remains the modality of choice for evaluation of suspected BE.
评估食管胶囊内镜(ECE)对胃食管反流病(GERD)患者巴雷特食管(BE)的诊断准确性。
检索关于评估ECE对GERD患者BE诊断准确性的盲法研究文献。进行荟萃分析以计算ECE诊断BE的合并敏感性和特异性。还根据所使用的参考标准进行亚组分析。
9项研究共纳入618例患者,符合纳入标准。所有研究中ECE诊断BE的合并敏感性和特异性分别为77%和86%。以食管胃十二指肠镜检查(EGD)为参考标准时,ECE诊断BE的合并敏感性和特异性分别为78%和90%;以组织学确诊的肠化生(IM)为参考标准时,合并敏感性和特异性分别为78%和73%。各研究间敏感性结果的统计学异质性不明显(P = 0.270,I² = 19),但特异性结果存在异质性(P < 0.001,I² = 74)。没有发表偏倚的证据。发现ECE安全且患者接受度高。
食管胶囊内镜对GERD患者BE的诊断具有中等的敏感性和特异性。EGD仍然是评估疑似BE的首选方式。