All-polyethylene compared with metal-backed tibial components in total knee arthroplasty at ten years. A prospective, randomized controlled trial.

作者信息

Bettinson Karen A, Pinder Ian M, Moran Chris G, Weir David J, Lingard Elizabeth A

机构信息

Department of Orthopaedics, Freeman Hospital, Newcastle upon Tyne NE7 7DN, United Kingdom.

出版信息

J Bone Joint Surg Am. 2009 Jul;91(7):1587-94. doi: 10.2106/JBJS.G.01427.

Abstract

BACKGROUND

Several studies have described equivalent performance on radiostereometric analysis at two years for metal-backed compared with all-polyethylene stemmed tibial implants. The purpose of this study was to determine the ten-year survivorship results of these two designs from a large randomized controlled trial.

METHODS

Patients who were fifty years old or more, with no history of infection, and were undergoing primary total knee arthroplasty were randomized at the time of surgery to receive either an all-polyethylene or a metal-backed tibial component. Patients were assessed preoperatively and at one, three, five, eight, and ten years postoperatively. All assessments included a clinical history, a physical examination, and a radiographic evaluation. A total of 510 consecutive patients (566 knees) were recruited from August 1993 to January 1997. The mean age of the patients at the time of the index arthroplasty was 69.3 years, and 299 (59%) were women. The primary diagnosis was osteoarthritis for 458 knees (80.9%) and rheumatoid arthritis for 108 knees (19.1%).

RESULTS

Two hundred and ninety-three patients returned for the ten-year follow-up evaluation. A total of twenty-eight knees had been revised. Ten-year survivorship, with revision for any reason (or the time at which patients were documented as requiring revision but were unfit for surgery) as the end point, was 94.5% (95% confidence interval, 90.4% to 96.8%) for the all-polyethylene design and 96% (95% confidence interval, 92.6% to 97.8%) for the metal-backed design. Ten-year survivorship, with aseptic failure as the end point, was 97% (95% confidence interval, 93.3% to 98.7%) for the all-polyethylene design and 96.8% (95% confidence interval, 93.6% to 98.4%) for the metal-backed design. On the basis of the numbers available at ten years, there was no significant difference in survivorship between the two designs (p > 0.05).

CONCLUSIONS

The long-term results demonstrate excellent survivorship, with revision as the end point, for both the metal-backed and the all-polyethylene tibial component designs with no differences noted between the two.

摘要

背景

多项研究表明,在两年的放射立体测量分析中,金属背衬型与全聚乙烯柄型胫骨植入物的性能相当。本研究的目的是通过一项大型随机对照试验确定这两种设计的十年生存率结果。

方法

年龄在50岁及以上、无感染史且正在接受初次全膝关节置换术的患者在手术时被随机分组,分别接受全聚乙烯或金属背衬的胫骨组件。在术前以及术后1年、3年、5年、8年和10年对患者进行评估。所有评估均包括临床病史、体格检查和影像学评估。从1993年8月至1997年1月共招募了510例连续患者(566个膝关节)。初次置换时患者的平均年龄为69.3岁,其中299例(59%)为女性。主要诊断为骨关节炎的膝关节有458个(80.9%),类风湿关节炎的膝关节有108个(19.1%)。

结果

293例患者返回进行十年随访评估。共有28个膝关节进行了翻修。以任何原因翻修(或记录为需要翻修但不适合手术的时间)为终点,全聚乙烯设计的十年生存率为94.5%(95%置信区间,90.4%至96.8%),金属背衬设计的十年生存率为96%(95%置信区间,92.6%至97.8%)。以无菌性失败为终点,全聚乙烯设计的十年生存率为97%(95%置信区间,93.3%至9