Suppr超能文献

磁共振成像引导下真空辅助乳腺活检:针对准确性的模型及患者评估

MRI-guided vacuum-assisted breast biopsy: a phantom and patient evaluation of targeting accuracy.

作者信息

El Khouli Riham H, Macura Katarzyna J, Barker Peter B, Elkady Laila M, Jacobs Michael A, Vogel-Claussen Jens, Bluemke David A

机构信息

Department of Radiology and Imaging Sciences, National Institutes of Health, Bethesda, Maryland 20892, USA.

出版信息

J Magn Reson Imaging. 2009 Aug;30(2):424-9. doi: 10.1002/jmri.21831.

Abstract

PURPOSE

To determine the spatial localization errors of magnetic resonance imaging (MRI)-guided core biopsy for breast lesions using the handheld vacuum-assisted core biopsy device in phantoms and patients.

MATERIALS AND METHODS

Biopsies were done using a 10-gauge handheld vacuum-assisted core biopsy system (Vacora, Bard, AZ, USA) on a 1.5T MRI scanner (Philips Achieva, Best, The Netherlands). A standardized biopsy localization protocol was followed by trained operators for multiplanar planning of the biopsy on a separate workstation. Biopsy localization errors were determined as the distance from needle tip to center of the target in three dimensions.

RESULTS

Twenty MRI-guided biopsies of phantoms were performed by three different operators. The biopsy target mean size was 6.8 +/- 0.6 mm. The overall mean three-dimensional (3D) biopsy targeting error was 4.4 +/- 2.9 mm. Thirty-two MRI breast biopsies performed in 22 patients were reviewed. The lesion mean size was 10.5 +/- 9.4 mm. The overall mean 3D localization error was 5.7 +/- 3.0 mm. No significant differences between phantom and patients biopsy errors were found (P > 0.5).

CONCLUSION

MRI-guided handheld vacuum-assisted core biopsy device shows good targeting accuracy and should allow localization of lesions to within approximately 5 to 6 mm.

摘要

目的

使用手持式真空辅助活检装置,在体模和患者中确定磁共振成像(MRI)引导下乳腺病变粗针活检的空间定位误差。

材料与方法

在1.5T MRI扫描仪(飞利浦Achieva,荷兰贝斯特)上,使用10号手持式真空辅助活检系统(Vacora,巴德公司,美国亚利桑那州)进行活检。经过培训的操作人员在单独的工作站上遵循标准化的活检定位方案,对活检进行多平面规划。活检定位误差确定为针尖到目标中心在三维空间中的距离。

结果

由三名不同的操作人员对体模进行了20次MRI引导下的活检。活检目标的平均大小为6.8±0.6毫米。总体平均三维(3D)活检靶向误差为4.4±2.9毫米。回顾了对22例患者进行的32次MRI引导下的乳腺活检。病变的平均大小为10.5±9.4毫米。总体平均3D定位误差为5.7±3.0毫米。在体模和患者活检误差之间未发现显著差异(P>0.5)。

结论

MRI引导的手持式真空辅助活检装置显示出良好的靶向准确性,应能将病变定位在约5至6毫米范围内。

相似文献

本文引用的文献

7
Magnetic resonance imaging of the breast prior to biopsy.活检前的乳腺磁共振成像。
JAMA. 2004 Dec 8;292(22):2735-42. doi: 10.1001/jama.292.22.2735.

文献检索

告别复杂PubMed语法,用中文像聊天一样搜索,搜遍4000万医学文献。AI智能推荐,让科研检索更轻松。

立即免费搜索

文件翻译

保留排版,准确专业,支持PDF/Word/PPT等文件格式,支持 12+语言互译。

免费翻译文档

深度研究

AI帮你快速写综述,25分钟生成高质量综述,智能提取关键信息,辅助科研写作。

立即免费体验