比较两种不同准分子激光行近视性光折射性角膜切削术后的地形消融区。

Comparison of the topographic ablation zone after photorefractive keratectomy for myopia using two different excimer lasers.

机构信息

Department of Ophthalmology, Ajou, Korea.

出版信息

Eye (Lond). 2010 Apr;24(4):553-7. doi: 10.1038/eye.2009.169. Epub 2009 Jul 24.

Abstract

PURPOSE

To compare the topographic features of eyes treated with photorefractive keratectomy (PRK) for myopia using two different excimer lasers.

METHODS

A total of 65 eyes in 39 patients treated with PRK (6.0-mm optical zone) using Technolas 217C and VISX S4 excimer lasers were evaluated retrospectively to determine the size of the topographic ablation zone.

RESULTS

The zones ablated using the VISX S4 had shorter diameters in both axes (-0.89+/-0.73, -1.59+/-0.49 mm; both P=0.00), whereas those ablated using the Technolas 217C had a longer diameter in the major axis (0.96+/-0.63 mm; P=0.00) and a shorter diameter in the minor axis (-0.39+/-0.59 mm; P=0.00). The theoretical ablated zone was a circle with a diameter of 6.0 mm. The Technolas 217C group tended to have oval cuts in comparison with the VISX S4 group, and the difference between the programmed (6.0 mm) and topographic diameters was significant in both groups.

CONCLUSIONS

There was a difference between the programmed and postoperative topographic diameters of the ablation zone. The postoperative ablation zone differed in shape and size according to the type of excimer laser.

摘要

目的

比较两种不同准分子激光治疗近视时行光折射性角膜切削术(PRK)的眼地形特征。

方法

回顾性分析 39 例(65 只眼)患者接受 PRK(6.0mm 光学区)治疗的资料,分别使用Technolas 217C 和 VISX S4 准分子激光,确定地形消融区的大小。

结果

VISX S4 组在两个轴上的消融区直径较小(-0.89+/-0.73,-1.59+/-0.49mm;均 P=0.00),而 Technolas 217C 组在长轴上的直径较大(0.96+/-0.63mm;P=0.00),在短轴上的直径较小(-0.39+/-0.59mm;P=0.00)。理论上消融区为直径 6.0mm 的圆。与 VISX S4 组相比,Technolas 217C 组的切割更倾向于椭圆形,两组间计划(6.0mm)与地形直径之间的差异均有统计学意义。

结论

消融区的计划和术后地形直径之间存在差异。根据准分子激光的类型,术后消融区的形状和大小存在差异。