Suppr超能文献

使用Amplatzer卵圆孔未闭封堵器进行经皮卵圆孔未闭封堵术:6个月随访时残余分流的预测因素

Percutaneous PFO closure with Amplatzer PFO occluder: predictors of residual shunts at 6 months follow-up.

作者信息

Greutmann Matthias, Greutmann-Yantiri Mehtap, Kretschmar Oliver, Senn Oliver, Roffi Marco, Jenni Rolf, Luescher Thomas F, Eberli Franz R

机构信息

Department of Cardiology, Toronto General Hospital, Toronto, Ontario, Canada.

出版信息

Congenit Heart Dis. 2009 Jul-Aug;4(4):252-7. doi: 10.1111/j.1747-0803.2009.00302.x.

Abstract

OBJECTIVE

The objective of this study was to assess predictors of residual shunts after percutaneous patent foramen ovale (PFO) closure with Amplatzer PFO occluder (AGA Medical Corporation, Golden Valley, MN, USA).

METHODS

All percutaneous PFO closures, using Amplatzer PFO occluder performed at a tertiary center between May 2002 and August 2006, were reviewed. Follow-up, including saline contrast transesophageal echocardiography, was performed in all patients 6 months after the intervention.

PATIENTS

A total of 135 procedures were performed. Mean age of the patients was 51 years. The indication for PFO closure was an ischemic cerebrovascular event in 92%, paradoxical systemic embolism in 4%, and a diving accident in 4%. Recurrent events prior to PFO closure were noted in 34%. A concomitant atrial septal aneurysm was present in 61%.

RESULTS

At 6 months follow-up, a residual shunt was detected in 26 patients (19%). Residual shunts were more common in patients with an atrial septal aneurysm (27 vs. 8%, P= .01) and in patients treated with a 35-mm compared with a 25-mm device (39 vs. 15%, P= .01). A concomitant atrial septal aneurysm remained independently associated with residual shunts when controlled for body mass index, gender, age, atrial dimensions, and presence of a Chiari network (odds ratio 4.1, 95% confidence intervals 1.1-15.0).

CONCLUSION

The presence of atrial septal aneurysms in patients undergoing percutaneous PFO closure with an Amplatzer PFO occluder significantly increases the rate of residual shunts at 6 months follow-up, even if 35-mm devices are used.

摘要

目的

本研究的目的是评估使用美国明尼苏达州黄金谷AGA医疗公司生产的Amplatzer卵圆孔未闭(PFO)封堵器经皮封堵PFO后残余分流的预测因素。

方法

回顾了2002年5月至2006年8月在一家三级中心使用Amplatzer PFO封堵器进行的所有经皮PFO封堵术。所有患者在干预后6个月进行随访,包括盐水对比经食管超声心动图检查。

患者

共进行了135例手术。患者的平均年龄为51岁。PFO封堵的适应证为缺血性脑血管事件占92%,反常性全身栓塞占4%,潜水事故占4%。34%的患者在PFO封堵术前有复发性事件。61%的患者伴有房间隔瘤。

结果

在6个月的随访中,26例患者(19%)检测到残余分流。残余分流在伴有房间隔瘤的患者中更常见(27%对8%,P = 0.01),并且与使用25mm装置相比,使用35mm装置治疗的患者中更常见(39%对15%,P = 0.01)。在控制体重指数、性别、年龄、心房大小和奇静脉网的存在后,伴有房间隔瘤仍与残余分流独立相关(比值比4.1,95%置信区间为[1.1,15.0])。

结论

使用Amplatzer PFO封堵器经皮封堵PFO的患者中,即使使用35mm装置,房间隔瘤的存在在6个月随访时也会显著增加残余分流的发生率。