Evidence for daptomycin Etest lot-related MIC elevations for Staphylococcus aureus.

机构信息

Laboratory Specialists Inc., Westlake, OH 44145, USA.

出版信息

Diagn Microbiol Infect Dis. 2009 Nov;65(3):306-11. doi: 10.1016/j.diagmicrobio.2009.07.012. Epub 2009 Sep 16.

Abstract

MIC testing was performed simultaneously by Etest and broth microdilution (BMD) on 587 Staphylococcus aureus isolates submitted by local laboratories to a reference laboratory for confirmatory testing (May 2005 to July 2008). Testing bias was assessed for Etest to BMD MIC ratios. Categoric and essential agreement, very major (BMD nonsusceptible, Etest susceptible), and major (BMD susceptible, Etest nonsusceptible) errors were evaluated. Agar and broth calcium concentrations were consistent with current Clinical and Laboratory Standards Institute and manufacturer recommendations. There was a consistent bias for higher Etest MIC values compared with BMD. Ratios ranged from 0.25 to 4 (average 1.3), with substantial variability noted among the 8 different Etest lots tested. Overall, 6% of all ratios were >2.0. Categoric agreement and essential agreement among the 8 Etest lots ranged from 73% to 96% and 74% to 100%, respectively; very major errors ranged from 3% to 9%, and major errors ranged from 6% to 35%. However, most of the discrepancies were limited to 3 Etest lots.

摘要

2005 年 5 月至 2008 年 7 月,当地实验室向参考实验室提交了 587 株金黄色葡萄球菌分离株进行确认检测,同时使用 Etest 和肉汤微量稀释法(BMD)进行 MIC 检测。评估了 Etest 与 BMD MIC 比值的检测偏倚。评估了分类和基本一致性、非常重大(BMD 耐药,Etest 敏感)和重大(BMD 敏感,Etest 耐药)错误。琼脂和肉汤中的钙浓度与当前的临床和实验室标准协会和制造商建议一致。与 BMD 相比,Etest 的 MIC 值存在一致的偏高趋势。比值范围从 0.25 到 4(平均为 1.3),在 8 个不同的 Etest 批次中存在显著的差异。总体而言,超过 2.0 的比值占所有比值的 6%。8 个 Etest 批次之间的分类和基本一致性分别为 73%至 96%和 74%至 100%;非常重大错误的范围为 3%至 9%,重大错误的范围为 6%至 35%。然而,大多数差异仅限于 3 个 Etest 批次。

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验