Mitamura Yoshinori, Takahashi Sayaka, Kano Kentaro, Okamoto Eiji, Murabayashi Shun, Nishimura Ikuya, Higuchi Taka-Aki
Department of Human Sciences and Informatics, School of Biological Science and Engineering, Tokai University, Sapporo, Japan.
Artif Organs. 2009 Sep;33(9):770-3. doi: 10.1111/j.1525-1594.2009.00899.x.
A magnetic fluid (MF) for a rotary blood pump seal enables mechanical contact-free rotation of the shaft and, hence, has excellent durability. The performance of a MF seal, however, has been reported to decrease in liquids. We have developed a MF seal that has a "shield" mechanism and a new MF with a higher magnetization of 47.9 kA/m. The sealing performance of the MF seal installed in a rotary blood pump was studied. Under the condition of continuous flow, the MF seal remained in perfect condition against a pressure of 298 mm Hg (pump flow rate: 3.96 L/min). The seal was also perfect against a pressure of 170 mm Hg in a continuous flow of 3.9 L/min for 275 days. We have developed a MF seal that works in liquid against clinically used pressures. The MF seal is promising as a shaft seal for rotary blood pumps.