Accuracy and reproducibility of stent-strut thickness determined by optical coherence tomography.

作者信息

Terashima Mitsuyasu, Rathore Sudhir, Suzuki Yasuyoshi, Nakayama Yoshiaki, Kaneda Hideaki, Nasu Kenya, Habara Maoto, Katoh Osamu, Suzuki Takahiko

机构信息

Division of Cardiovascular Medicine, Toyohashi Heart Center, 21-1 Gobudori, Oyama-cho, Toyohashi, 441-8530, Japan.

出版信息

J Invasive Cardiol. 2009 Nov;21(11):602-5.

Abstract

BACKGROUND

Optical coherence tomography (OCT) has been increasingly used to evaluate stent apposition following implantation. Since stent struts are visualized as linear structures with strong surface reflection and typical dorsal shadowing, apposition of struts is evaluated by measuring the distance between the strut surface reflection and adjacent vessel surface in consideration of strut thickness. However, there are no data available to validate the measurements of strut thickness by OCT. The aim of this in vitro study is to validate the accuracy of OCT measurement of stent-strut thickness of different commercially available stents in evaluating stent apposition.

METHODS

We performed the in vitro study after implantation of 5 commonly used stents in a phantom model artery. Stent-strut thickness was measured by a commercially available OCT system and was compared to the manufacturers' nominal strut-thickness data for each stent. Intra- and interobserver variability were also assessed.

RESULTS

A total of 239 stent struts were evaluated. The differences in stent-strut measurements as compared to the manufacturers' nominal strut thickness data were low. The intra- and interobserver measurement differences were low (6 +/- 7 microm, and 6 +/- 7 microm, respectively), with high correlation coefficients (r = 0.957 and r = 0.957, respectively; p < 0.0001).

CONCLUSIONS

This in vitro study demonstrates that OCT analysis measuring stent-strut thickness provides accurate data with high reproducibility, suggesting that assessment of stent-strut apposition using OCT is feasible.

摘要

背景

光学相干断层扫描(OCT)已越来越多地用于评估支架植入后的贴壁情况。由于支架小梁被可视化为具有强烈表面反射和典型背侧阴影的线性结构,因此在考虑小梁厚度的情况下,通过测量小梁表面反射与相邻血管表面之间的距离来评估小梁的贴壁情况。然而,尚无可用数据来验证通过OCT测量小梁厚度的准确性。本体外研究的目的是验证OCT测量不同市售支架的支架小梁厚度在评估支架贴壁情况时的准确性。

方法

我们在模拟模型动脉中植入5种常用支架后进行了体外研究。使用市售的OCT系统测量支架小梁厚度,并将其与每个支架的制造商标称小梁厚度数据进行比较。还评估了观察者内和观察者间的变异性。

结果

共评估了239个支架小梁。与制造商标称的小梁厚度数据相比,支架小梁测量的差异较小。观察者内和观察者间的测量差异较小(分别为6±7微米和6±7微米),相关系数较高(分别为r = 0.957和r = 0.957;p < 0.0001)。

结论

本体外研究表明,OCT分析测量支架小梁厚度可提供具有高重现性的准确数据,这表明使用OCT评估支架小梁贴壁情况是可行的。