Anesth Analg. 2010 Jan 1;110(1):94-100. doi: 10.1213/ANE.0b013e3181c22021. Epub 2009 Nov 12.
Oxygen supplies capable of supporting transtracheal jet ventilators can be lifesaving. There is not much information about which oxygen sources (readily available inside and outside operating rooms) have sufficient driving pressure for transtracheal jet ventilation.
We measured driving pressures (upstream or residual oxygen pressure) in a specially designed jet ventilation test system with a 2.25-mm (14-gauge) i.v. catheter. High-pressure oxygen sources evaluated included wall-mounted (Puritan, Allied Health, Precision, and Datex-Ohmeda) and anesthesia machine auxiliary oxygen flowmeters and oxygen flush valves from anesthesia machines (Draeger Narkomed 2B, Narkomed 4, Datex-Ohmeda Excel, and Datex-Ohmeda Modulus).
All 4 types of wall-mounted oxygen flowmeters, opened past their highest scale settings (15 L/min), delivered sufficient working pressures (range, 103-282 kPa; 16-41 psi). Working pressures from auxiliary oxygen flowmeters mounted on Datex-Ohmeda machines were adequate to support jet ventilation (range, 189-248 kPa; 27-36 psi), whereas those on tested Draeger machines did not supply sufficient pressure for jet ventilation: Narkomed 2B, 14-28 kPa (2-4 psi); Narkomed 4, 24-28 kPa (3-4 psi). Working pressures delivered by oxygen flush valves on tested Draeger machines were adequate to support jet ventilation, ranging from 96 to 117 kPa (14-17 psi), whereas pressures generated by tested Datex-Ohmeda flush valves were not (ranging from 50 to 62 kPa, 7-9 psi).
Oxygen sources other than dedicated jet ventilator connectors to high-pressure pipeline oxygen may supply adequate working pressure, but each type of oxygen source needs testing to ensure that it supplies adequate working pressure.
能够提供支持经气管喷射通气的氧气供应可能是救命的。关于哪些氧气源(手术室内外均可轻易获得)具有足够的经气管喷射通气驱动压力,相关信息并不多。
我们使用专门设计的喷射通气测试系统,通过一根 2.25 毫米(14 号)静脉导管测量驱动压力(上游或残留氧气压力)。评估的高压氧气源包括壁挂式(Puritan、Allied Health、Precision 和 Datex-Ohmeda)和麻醉机辅助氧气流量计以及麻醉机的氧气冲洗阀(Draeger Narkomed 2B、Narkomed 4、Datex-Ohmeda Excel 和 Datex-Ohmeda Modulus)。
所有 4 种壁挂式氧气流量计,在超过其最高刻度设置(15 L/min)时,均能提供足够的工作压力(范围为 103-282 kPa;16-41 psi)。Datex-Ohmeda 机器上的辅助氧气流量计提供的工作压力足以支持喷射通气(范围为 189-248 kPa;27-36 psi),而在测试的 Draeger 机器上的则不能提供足够的喷射通气压力:Narkomed 2B,14-28 kPa(2-4 psi);Narkomed 4,24-28 kPa(3-4 psi)。测试的 Draeger 机器上的氧气冲洗阀提供的工作压力足以支持喷射通气,范围为 96-117 kPa(14-17 psi),而测试的 Datex-Ohmeda 冲洗阀产生的压力则不够(范围为 50-62 kPa,7-9 psi)。
除了专用的经气管喷射通气连接器到高压管道氧气之外的氧气源可能提供足够的工作压力,但每种氧气源都需要进行测试,以确保其提供足够的工作压力。