Bergeron Denis E, Zimmerman Brian E, Cessna Jeffrey T
Physics Laboratory, National Institute of Standards and Technology, Gaithersburg, MD 20899, USA.
Appl Radiat Isot. 2010 Jul-Aug;68(7-8):1367-70; discussion 1370. doi: 10.1016/j.apradiso.2009.11.005. Epub 2009 Nov 10.
Ra-223 is a bone-seeking alpha emitter currently being evaluated as a radiopharmaceutical. Concurrent with the primary standardization, NIST established that calibration factors currently used for radionuclide calibrators in the clinical setting give readings 5.7-8.7% higher than the NIST calibrated activity. This work describes the determination of calibration factors specific to dose vials and syringes. Using the calibration factors derived with standard ampoules to measure syringe activities can give readings up to 3.6% too high.