A new method that utilizes 1-(2-aminoethyl)-3-phenylurea-modified silica gel as a solid-phase extractant has been developed for preconcentration of trace Sc(III) prior to the measurement by inductively coupled plasma atomic emission spectrometry (ICP-AES). Experimental conditions for effective adsorption of trace level of Sc(III) were optimized using batch and column procedures in detail. The optimum pH value for the separation of Sc(III) on the new sorbent was 4 and complete elution of Sc(III) from the sorbent surface was carried out using 1.0mL of 0.1molL(-1) HCl. Common coexisting ions did not interfere with the separation and determination of the analyte. The maximum static adsorption capacity of the sorbent at optimum conditions was found to be 32.5mgg(-1) while the time of 95% adsorption was less than 2min. The detection limit of present method was found to be 0.091microgg(-1), and the relative standard deviation (RSD) was lower than 3.0% (n=8). The method was successfully applied for the preconcentration of trace Sc(III) in the environmental samples with satisfactory results.
一种新的方法,利用 1-(2-氨乙基)-3-苯基脲修饰硅胶作为固相萃取剂,已被开发用于电感耦合等离子体原子发射光谱法(ICP-AES)测量前痕量 Sc(III)的预浓缩。通过批处理和柱处理详细优化了有效吸附痕量 Sc(III)的实验条件。在新的吸附剂上分离 Sc(III)的最佳 pH 值为 4,使用 1.0mL 0.1molL(-1)HCl 可从吸附剂表面完全洗脱 Sc(III)。常见共存离子不干扰分析物的分离和测定。在最佳条件下,吸附剂的最大静态吸附容量为 32.5mgg(-1),而 95%吸附的时间小于 2min。本方法的检测限为 0.091microgg(-1),相对标准偏差(RSD)低于 3.0%(n=8)。该方法成功地应用于环境样品中痕量 Sc(III)的预浓缩,结果令人满意。