Suppr超能文献

长寿命放射性核素的半衰期测量——新数据分析与系统效应

Half-life measurements of long-lived radionuclides--new data analysis and systematic effects.

作者信息

Schrader H

机构信息

Physikalisch-Technische Bundesanstalt (PTB), Braunschweig, Germany.

出版信息

Appl Radiat Isot. 2010 Jul-Aug;68(7-8):1583-90. doi: 10.1016/j.apradiso.2009.11.033. Epub 2009 Nov 18.

Abstract

Half-lives of long-lived radionuclides, for example, (85)Kr, (90)Sr, (108m)Ag, (133)Ba, (137)Cs, (152)Eu and (154)Eu, were measured at the Physikalisch-Technische Bundesanstalt (PTB) by ionization chamber measuring systems, for some nuclides, during a period of about 30 years. In particular, the relative residuals of fitted raw data from current measurements by a modified Townsend balance (compensation method) showed small periodic yearly fluctuations. The interpretation of these fluctuations is discussed. To solve problems of instrument instabilities (fluctuations), a new method of data analysis is applied using least-squares fits of combinations of the current ratios of the various radionuclide sources with corresponding data points measured each at nearly the same time. The data are corrected for instabilities similarly to when using current ratios with a (226)Ra reference source. The new method of data analysis excludes possible systematic effects due to disequilibrium of the (226)Ra decay chain. The limits of the method are discussed including the elimination of outliers due to systematic errors. An uncertainty budget is presented with the corresponding uncertainty components. Relative residuals of the fits in the best cases of the order of +/-5x10(-4) have been achieved. The results comprise data until the end of 2008. They agree within the limits of their relative uncertainties, typically better than 1.5x10(-3), with those published earlier by PTB authors.

摘要

例如,长寿命放射性核素氪-85、锶-90、银-108m、钡-133、铯-137、铕-152和铕-154的半衰期,在约30年的时间里,由德国物理技术联邦研究所(PTB)使用电离室测量系统进行了测量。特别是,通过改进的汤森平衡(补偿法)对当前测量的原始数据进行拟合得到的相对残差显示出每年有小的周期性波动。本文讨论了这些波动的解释。为了解决仪器不稳定性(波动)问题,应用了一种新的数据分析方法,即对各种放射性核素源的电流比与几乎在同一时间测量的相应数据点的组合进行最小二乘拟合。数据的不稳定性校正方法与使用镭-226参考源的电流比时类似。这种新的数据分析方法排除了由于镭-226衰变链不平衡可能产生的系统效应。讨论了该方法的局限性,包括消除由于系统误差导致的异常值。给出了包含相应不确定度分量的不确定度预算。在最佳情况下,拟合的相对残差达到了±5×10⁻⁴的量级。结果包含截至2008年底的数据。它们在相对不确定度范围内(通常优于1.5×10⁻³)与PTB作者早期发表的结果一致。

相似文献

1
Half-life measurements of long-lived radionuclides--new data analysis and systematic effects.
Appl Radiat Isot. 2010 Jul-Aug;68(7-8):1583-90. doi: 10.1016/j.apradiso.2009.11.033. Epub 2009 Nov 18.
2
Seasonal variations of decay rate measurement data and their interpretation.
Appl Radiat Isot. 2016 Aug;114:202-13. doi: 10.1016/j.apradiso.2016.05.001. Epub 2016 May 16.
3
Half-life measurements with ionization chambers--a study of systematic effects and results.
Appl Radiat Isot. 2004 Feb-Apr;60(2-4):317-23. doi: 10.1016/j.apradiso.2003.11.039.
4
Concerning the time dependence of the decay rate of 137Cs.
Appl Radiat Isot. 2013 Apr;74:50-5. doi: 10.1016/j.apradiso.2012.12.010. Epub 2012 Dec 31.
5
Calibration of a radionuclide calibrator system as a Bulgarian standard for activity.
Appl Radiat Isot. 2008 Jun-Jul;66(6-7):965-71. doi: 10.1016/j.apradiso.2008.02.042. Epub 2008 Feb 13.
6
Standardization of radium-223 by liquid scintillation counting.
Appl Radiat Isot. 2010 Jul-Aug;68(7-8):1523-8. doi: 10.1016/j.apradiso.2009.11.068. Epub 2009 Dec 2.
8
Comparison of ionization chamber efficiencies for activity measurements.
Appl Radiat Isot. 2004 Feb-Apr;60(2-4):369-78. doi: 10.1016/j.apradiso.2003.11.044.
10
Uncertainties in alanine/ESR dosimetry at the Physikalisch-Technische Bundesanstalt.
Phys Med Biol. 2006 Nov 7;51(21):5419-40. doi: 10.1088/0031-9155/51/21/003. Epub 2006 Oct 6.

引用本文的文献

1
Believable Statements of Uncertainty and Believable Science.
J Radioanal Nucl Chem. 2017 Feb;311(2):1019-1022. doi: 10.1007/s10967-016-4912-4. Epub 2016 Jun 21.

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验