Suppr超能文献

替代测量胫后神经体感诱发电位的前参考部位。

Alternative anterior reference sites for measuring posterior tibial nerve somatosensory evoked potentials.

机构信息

Department of Anesthesiology, Northwestern University Feinberg School of Medicine, Chicago, IL, USA.

出版信息

J Clin Monit Comput. 2010 Apr;24(2):107-11. doi: 10.1007/s10877-009-9216-6. Epub 2010 Jan 9.

Abstract

OBJECTIVE

The purpose of this study was to examine the utility and feasibility of using alternative anterior reference leads when measuring left posterior tibial nerve somatosensory evoked potentials (SEPs).

METHODS

With IRB approval, 12 patients were monitored using both traditional (FPz and C4') and alternative anterior (F3 and F4) reference leads during routine spine surgery with SEP monitoring. Recordings from the routine and novel electrode pairs were collected and analyzed.

RESULTS

All of the SEP amplitudes measured were of similar magnitude except for that of F3-F4, which was significantly lower (P < 0.001) than all of the other five lead combinations which were assessed (Cz'-FPz, C3'-C4', C3'-F4, Cz'-F3, and Cz'-F4). The latencies of the novel lead combinations (C3'-F4, Cz'-F3, Cz'-F4, and F3-F4) were similar to those of the "gold standards" (Cz'-FPz and C3'-C4') (pooled median, 45.6 ms with 25-75th percentiles, 44.0-47.8 ms, P = 0.308). The coefficients of variation (CV %) of the amplitudes were not statistically significantly different (P = 0.341).

CONCLUSIONS

The use of alternative frontal reference leads (F3 and F4) for left posterior tibial nerve SEP monitoring yields signals of equal quality and reproducibility compared to signals with standard (FPz and C4') referencing. These alternative leads may substitute for traditional referencing when placement of FPz or C4' is precluded by the location of surgery.

摘要

目的

本研究旨在探讨在测量左后胫神经体感诱发电位(SEP)时使用替代额部参考导联的实用性和可行性。

方法

在机构审查委员会批准下,12 例患者在接受脊柱手术伴 SEP 监测时,使用传统(FPz 和 C4')和替代额部(F3 和 F4)参考导联进行监测。收集和分析常规和新型电极对的记录。

结果

除 F3-F4 外,所有测量的 SEP 幅度均相似,F3-F4 明显较低(P < 0.001),低于评估的其他五个导联组合(Cz'-FPz、C3'-C4'、C3'-F4、Cz'-F3 和 Cz'-F4)。新型导联组合(C3'-F4、Cz'-F3、Cz'-F4 和 F3-F4)的潜伏期与“金标准”(Cz'-FPz 和 C3'-C4')相似(中位数,45.6ms,25-75%分位数为 44.0-47.8ms,P = 0.308)。幅度的变异系数(CV%)无统计学差异(P = 0.341)。

结论

与标准参考(FPz 和 C4')相比,使用替代额部参考导联(F3 和 F4)进行左后胫神经 SEP 监测可产生质量和可重复性相当的信号。当手术部位限制 FPz 或 C4'的放置时,这些替代导联可以替代传统参考导联。