DeGrazio Frances, Runkle Joseph, Smythe Jeff, Miller Amy
West Pharmaceutical Services, 101 Gordon Drive, Lionville, PA 19341, USA.
PDA J Pharm Sci Technol. 2009 Jul-Aug;63(4):360-7.
A study was designed to evaluate a new syringe system developed specifically for sensitive pharmaceutical and biopharmaceutical drug products. An in-depth review of the extractables portion of this program is presented and includes data generated by liquid chromatography-mass spectroscopy, gas chromatography-mass spectroscopy, ion chromatography, and inductively coupled plasma of the Crystal Zenith barrel before and after sterilization. One leachable was identified after e-beam irradiation processing. This leachable, identified as an organic acid species, was then evaluated from a safety assessment viewpoint. The details of this program and associated results are explained in this report.
一项研究旨在评估一种专门为敏感的药品和生物制药产品开发的新型注射器系统。本文对该项目可提取物部分进行了深入综述,包括灭菌前后Crystal Zenith针筒通过液相色谱-质谱联用、气相色谱-质谱联用、离子色谱和电感耦合等离子体质谱法生成的数据。电子束辐照处理后鉴定出一种可浸出物。该可浸出物被鉴定为一种有机酸,随后从安全评估的角度进行了评估。本报告解释了该项目的详细情况及相关结果。