1 个月后成功植入卵圆孔未闭封堵器后发生血栓形成、骨折和经皮取出。

Thrombosis, fracture, and percutaneous removal of a patent foramen ovale closure device 1 month after successful deployment.

机构信息

II Medizinische Klinik für Kardiologie/Angiologie, Johannes Gutenberg University, Mainz, Germany.

出版信息

Catheter Cardiovasc Interv. 2010 Apr 1;75(5):778-81. doi: 10.1002/ccd.22320.

Abstract

Several different devices have been developed for the percutaneous closure of interatrial defects and patent foramen ovale. Although the implantation of these devices is both safe and effective, a number of complications, both in the early and the late follow-up, may occur. We describe a case of device fracture manifested early (1 month after implantation) with the formation of massive thrombosis on the right atrial disc. The patient was treated with anticoagulants and the device was percutaneously retrieved. Our images allowed early noninvasive therapy and emphasize the need for echocardiographic follow-up early after implantation.

摘要

已经开发出几种不同的设备用于经皮关闭房间隔缺损和卵圆孔未闭。尽管这些设备的植入既安全又有效,但在早期和晚期随访中可能会发生一些并发症。我们描述了一例早期(植入后 1 个月)出现设备断裂的病例,表现为右心房盘上形成大量血栓。患者接受抗凝治疗,并经皮取出设备。我们的图像实现了早期的无创治疗,并强调了在植入后早期进行超声心动图随访的必要性。