连续波多普勒和多巴酚丁胺负荷超声心动图对圣犹达医疗 Regent 人工瓣膜的评估。

Assessment of the St. Jude Medical Regent prosthetic valve by continuous-wave Doppler and dobutamine stress echocardiography.

机构信息

Department of Cardiovascular Surgery, Sekino Hospital, Nihon University School of Medicine, Tokyo, Japan.

出版信息

Ann Thorac Surg. 2010 Jan;89(1):87-92. doi: 10.1016/j.athoracsur.2009.09.013.

Abstract

BACKGROUND

The St. Jude Medical (SJM) Regent prosthetic valve (St. Paul, MN), a recently developed mechanical valve, is an improvement on the conventional SJM valve, having a wider valve area than the SJM HP valve. We evaluated this mechanical valve by Doppler echocardiography and dobutamine stress echocardiography (DSE).

METHODS

The functions of the SJM Regent valve were evaluated by continuous-wave Doppler echocardiography and DES in 58 cases of aortic valve replacement during a stable postoperative period.

RESULTS

The peak pressure gradient of the replaced valves sized 17, 19, 21, and 23 mm was 27.5 +/- 11.1, 20.0 +/- 9.8, 15.6 +/- 5.7, and 14.3 +/- 9.1, mm Hg respectively, and the effective orifice area index was 0.97 +/- 0.32, 1.01 +/- 0.29, 1.09 +/- 0.30, and 1.41 +/- 0.54 cm(2)/m(2), respectively, with prosthesis-patient mismatch (PPM) found in 1, 3, 2, and 0 cases for the 17-, 19-, 21- and 23-mm valves, respectively, with a total incidence of 10.3%. In 20 cases, the peak pressure gradient and the effective orifice area index were significantly increased during DSE compared with those at rest.

CONCLUSIONS

Although the PPM incidence was 6.6%, it was deemed from the data of DSE and clinical symptoms that there were no clinical issues for such cases of PPM in the early and intermediate phases after operation. Particularly, the effectiveness of the 17- and 19-mm valves in patients with a small aortic annulus was demonstrated, confirming the satisfactory functions of the SJM Regent prosthetic valve.

摘要

背景

圣犹达医疗(SJM)Regent 人工心脏瓣膜(明尼苏达州圣保罗)是最近开发的一种机械瓣膜,它是对传统 SJM 瓣膜的改进,其瓣膜面积比 SJM HP 瓣膜更大。我们通过多普勒超声心动图和多巴酚丁胺负荷超声心动图(DSE)对这种机械瓣膜进行了评估。

方法

在稳定的术后期间,通过连续波多普勒超声心动图和 DSE 对 58 例主动脉瓣置换术的 SJM Regent 瓣膜功能进行了评估。

结果

置换瓣膜尺寸为 17、19、21 和 23mm 的峰值压力梯度分别为 27.5±11.1、20.0±9.8、15.6±5.7 和 14.3±9.1mmHg,有效瓣口面积指数分别为 0.97±0.32、1.01±0.29、1.09±0.30 和 1.41±0.54cm2/m2,17、19、21 和 23mm 瓣膜的 PPM 分别为 1、3、2 和 0 例,总发生率为 10.3%。在 20 例患者中,与静息时相比,DSE 期间峰值压力梯度和有效瓣口面积指数显著增加。

结论

尽管 PPM 发生率为 6.6%,但从 DSE 和临床症状的数据来看,在术后早期和中期,这种 PPM 病例没有临床问题。特别是,对于主动脉瓣环较小的患者,17 和 19mm 瓣膜的有效性得到了证实,证实了 SJM Regent 人工心脏瓣膜的满意功能。