Department of Rheumatology, Ghent University Hospital 0K12-IB, De Pintelaan 185, B-900 Ghent, Belgium.
Clin Exp Rheumatol. 2009 Nov-Dec;27(6):971-6.
Although indirect immunofluorescence (IIF) is the most widely applied screening test for antinuclear antibodies (ANA), it lacks specificity for the identification of specific diseases or antigen reactivities. The aim of the present study was to validate an anti-extractable nuclear antigen (ENA) screening strategy encompassing a three-step cascade whereby an ELISA with pooled specific ENA is positioned between the IIF and the final anti-ENA identification.
Sera from 4 populations were tested for anti-ENA using an automated ELISA (EliA Symphony) and a line immunoassay (INNO-LIA ANA update).
At the manufacturer's cut-off, a 96% sensitivity (95% CI 94%-98%) and 96% specificity (95% CI 94%-98%) of EliA Symphony for anti-ENA was obtained in a consecutive selection of 328 IIF positive serum samples referred for ANA testing. In addition, a high sensitivity was demonstrated for anti-ENA reactivities in patients with SLE (99%, 95% CI 97%-101%) and SSc (100%), and for anti-ENA monoreactivities.
The EliA Symphony test was shown to be a sensitive second-line screening test for anti-ENA antibodies. In the context of a high clinical suspicion of connective tissue disease or autoreactivities not included in the EliA Symphony assay, third-line testing may be useful, even if the anti-ENA screening is negative.
虽然间接免疫荧光法(IIF)是最广泛应用的抗核抗体(ANA)筛选试验,但它缺乏对特定疾病或抗原反应性的特异性识别。本研究的目的是验证一种抗可提取核抗原(ENA)的筛选策略,该策略包括一个三步级联,其中 ELISA 与特定的 ENA 池一起定位在 IIF 和最终的抗 ENA 鉴定之间。
使用自动 ELISA(EliA Symphony)和线免疫分析(INNO-LIA ANA update)对 4 个人群的血清进行抗 ENA 检测。
在制造商的截止值下,EliA Symphony 在对 328 例连续选择的 IIF 阳性血清样本进行 ANA 检测时,对抗 ENA 的敏感性为 96%(95%CI 94%-98%),特异性为 96%(95%CI 94%-98%)。此外,该检测还显示出对 SLE(99%,95%CI 97%-101%)和 SSc(100%)患者的抗 ENA 反应性以及抗 ENA 单反应性具有较高的敏感性。
EliA Symphony 试验被证明是一种敏感的抗 ENA 抗体二线筛选试验。在高度怀疑结缔组织病或不在 EliA Symphony 检测范围内的自身反应性的情况下,即使抗 ENA 筛选呈阴性,三线检测可能也是有用的。