吸引管连接到吸引仪器上的微生物污染及预防方法。

Microbial contamination of suction tubes attached to suction instruments and preventive methods.

机构信息

Department of Pharmacy, Shunan Municipal Shinnanyo Citizen Hospital, Miyanomae, Shunan, Japan.

出版信息

Jpn J Infect Dis. 2010 Mar;63(2):124-7.

Abstract

We investigated the microbial contamination of suction tubes attached to wall-type suction instruments. Microbial contamination of suction tubes used for endoscopy or sputum suction in hospital wards was examined before and after their disinfection. In addition, disinfection and washing methods for suction tubes were evaluated. Suction tubes (n=33) before disinfection were contaminated with 10(2)-10(8) colony-forming units (cfu)/tube. The main contaminants were Pseudomonas aeruginosa, Acinetobacter baumannii, and Stenotrophomonas maltophilia. The suction tubes were disinfected with sodium hypochlorite (n=11) or hot water (n=11), or by an automatic tube cleaner (n=11). After 2-h immersion in 0.1% (1,000 ppm) sodium hypochlorite, 10(3)-10(7) cfu/tube of bacteria were detected in all 11 tubes examined. After washing in hot running water (65 degrees C), 10(3)-10(7) cfu/tube were detected in 3 of the 11 examined tubes. The bacteria detected in the suction tubes after disinfection with sodium hypochlorite or hot water were P. aeruginosa, A. baumannii, and S. maltophilia. On the other hand, after washing with warm water (40 degrees C) using the automatic tube cleaner, contamination was found to be <20 cfu/tube (lower detection limit, 20 cfu/tube) in all 11 tubes examined. These results suggest the usefulness of washing with automatic tube cleaners.

摘要

我们调查了附在墙壁式吸引仪器上的吸引管的微生物污染情况。在对病房内用于内镜或吸痰的吸引管进行消毒前后,对其微生物污染情况进行了检测。此外,还评估了吸引管的消毒和清洗方法。在消毒前,33 根吸引管中污染的细菌数量为 10(2)-10(8)cfu/管。主要污染物为铜绿假单胞菌、鲍曼不动杆菌和嗜麦芽窄食单胞菌。用次氯酸钠(n=11)或热水(n=11)或自动管清洗器(n=11)对吸引管进行消毒。在 0.1%(1000ppm)次氯酸钠中浸泡 2 小时后,在检查的 11 个管中均检测到 10(3)-10(7)cfu/管的细菌。在 65°C 的热水中冲洗后,在检查的 11 个管中有 3 个管中检测到 10(3)-10(7)cfu/管。在次氯酸钠或热水消毒后的吸引管中检测到的细菌为铜绿假单胞菌、鲍曼不动杆菌和嗜麦芽窄食单胞菌。另一方面,在用自动管清洗器用温水(40°C)清洗后,在检查的 11 个管中均未检测到 20cfu/管以下的污染(检测下限 20cfu/管)。这些结果表明,使用自动管清洗器进行清洗是有效的。

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验