Suppr超能文献

[根测仪诊断水平根折的体外研究]

[An in vitro study of Root ZX root apex locator to diagnose horizontal root fracture].

作者信息

Zhang Ping

机构信息

Dept of Stomatology, The Third Hospital of Peking University, Beijing 100191, China.

出版信息

Hua Xi Kou Qiang Yi Xue Za Zhi. 2010 Feb;28(1):102-3.

Abstract

OBJECTIVE

To evaluate the accuracy of Root ZX root apex locator in detecting stimulated horizontal root fractures.

METHODS

A total of 20 single-canal maxillary anterior teeth were collected. All roots were simulated horizontal root fracture. Root ZX root apex locator measured the canal length in 3 stages: Broken ends of horizontal fractures without separation; broken ends of horizontal fractures with separation, but no impression material; broken ends of horizontal fractures with separation and impression material. The lengths measured by Root ZX root apex locator were compared with actual canal length.

RESULTS

To horizontal root fracture without separation or without impression material, Root ZX root apex locator couldn't detect the horizontal root fracture. To horizontal root fracture with separation and impression material, Root ZX root apex locator detected the horizontal root fracture within +/- 0.5 mm in 100%.

CONCLUSION

Root ZX root apex locator lacks of diagnostic value for horizontal root fractures without soft tissue ingrowth, but provides with preferable veracity for horizontal root fractures with soft tissue ingrowth.

摘要

目的

评估Root ZX根尖定位仪检测模拟水平根折的准确性。

方法

收集20颗单根管上颌前牙。所有牙根均模拟水平根折。Root ZX根尖定位仪在3个阶段测量根管长度:水平骨折断端未分离;水平骨折断端分离但无印模材料;水平骨折断端分离且有印模材料。将Root ZX根尖定位仪测量的长度与实际根管长度进行比较。

结果

对于未分离或无印模材料的水平根折,Root ZX根尖定位仪无法检测到水平根折。对于有分离且有印模材料的水平根折,Root ZX根尖定位仪100%能在±0.5mm范围内检测到水平根折。

结论

Root ZX根尖定位仪对无软组织长入的水平根折缺乏诊断价值,但对有软组织长入的水平根折具有较好的准确性。