Thomas Christoph, Ketelsen Dominik, Tsiflikas Ilias, Reimann Anja, Brodoefel Harald, Claussen Claus D, Heuschmid Martin
Department of Diagnostic and Interventional Radiology, University of Tübingen, Tübingen, Germany.
J Comput Assist Tomogr. 2010 Mar-Apr;34(2):309-15. doi: 10.1097/RCT.0b013e3181c9b9ff.
To assess dose and image quality of dual-energy (DE) mixed images in comparison to single-energy (SE) images.
A phantom containing iodine contrast inserts was scanned using SE and DE protocols. Dual-energy mixed images were reconstructed with varying composition ratios (ratio describing the contribution of 80 and 140 data in the mixed images). Image noise, iodine contrast, and contrast-to-noise ratios (CNRs) were assessed inside and outside the central field of view (FoV).
With the default composition ratio (0.3), noise and contrast were comparable between both protocols in the central FoV. Peripherally, DE image noise exceeded noise in SE images; CNR in the periphery was lower in the DE images. The highest CNR was found for a composition ratio of 0.5 to 0.9, exceeding the CNR of SE images.
Dual-energy mixed images offer an image quality comparable to SE images within the central FoV at comparable dose levels. In the peripheral FoV, image quality is decreased. By optimizing the composition ratio in the DE mixed images, higher CNRs than in the SE images can be achieved, leading to a dose reduction potential.
评估双能(DE)混合图像与单能(SE)图像相比的剂量和图像质量。
使用SE和DE协议对包含碘对比剂插入物的体模进行扫描。以不同的合成比(描述混合图像中80和140数据贡献的比率)重建双能混合图像。在中心视野(FoV)内外评估图像噪声、碘对比度和对比度噪声比(CNR)。
在默认合成比(0.3)下,两种协议在中心FoV中的噪声和对比度相当。在外围,DE图像噪声超过SE图像中的噪声;DE图像在外围的CNR较低。合成比为0.5至0.9时CNR最高,超过了SE图像的CNR。
双能混合图像在可比剂量水平下,在中心FoV内提供与SE图像相当的图像质量。在外围FoV中,图像质量下降。通过优化DE混合图像中的合成比,可以实现比SE图像更高的CNR,从而具有降低剂量的潜力。