Zhang Haizhu, Cui Changcong, Hu Dayi
Department of Cardiology, Anyang People's Hospital, Anyang, 457000, China.
J Huazhong Univ Sci Technolog Med Sci. 2010 Apr;30(2):155-8. doi: 10.1007/s11596-010-0204-3. Epub 2010 Apr 21.
The effects of electric currents applied during absolute refractory period (ARP) on the cardiac function of rabbits with heart failure due to myocardial infarction (MI), and the safety of this method were investigated. Thirty rabbits were randomly assigned equally to 3 groups: sham-operated group, LV-anterior wall cardiac contractility modulation (LV-CCM) group, and septum-CCM (S-CCM) group. A thoracotomy was performed on all the rabbits. Electric pulses were delivered during the ARP on the anterior wall of left ventricle in CCM group and in the septum in S-CCM group, respectively. The left ventricular systolic pressure (LVSP) and maximum positive left ventricular pressure change (+dp/dt(max)), heart rates, ventricular tachycardia, ventricular fibrillation were observed. It was found that, as compared with the baseline, LVSP, and +dp/dtmax were significantly increased, on average, by 15.2% and 19.5% in LV-CCM group (P<0.05), and by 8.5% and 10.8% in S-CCM group (P<0.05). LVEDP was significantly decreased and -dp/dt(max) increased both in LV-CCM group and S-CCM group (P<0.05). CCM had no effect on heart rate and induced no arrhythmia in short time. It is concluded that electric currents delivered during the ARP could significantly enhance the contractility of myocardium safely, suggesting that CCM stimulation is a novel potent method for contractility modulation.
研究了在绝对不应期(ARP)施加电流对心肌梗死(MI)所致心力衰竭家兔心功能的影响以及该方法的安全性。30只家兔随机均分为3组:假手术组、左室前壁心脏收缩力调制(LV-CCM)组和室间隔CCM(S-CCM)组。对所有家兔进行开胸手术。CCM组在ARP期间于左心室前壁施加电脉冲,S-CCM组在室间隔施加电脉冲。观察左心室收缩压(LVSP)、左心室最大压力上升速率(+dp/dt(max))、心率、室性心动过速、心室颤动情况。结果发现,与基线相比,LV-CCM组LVSP和+dp/dtmax平均分别显著升高15.2%和19.5%(P<0.05),S-CCM组分别升高8.5%和10.8%(P<0.05)。LV-CCM组和S-CCM组左心室舒张末期压力(LVEDP)均显著降低,-dp/dt(max)升高(P<0.05)。CCM对心率无影响,短时间内未诱发心律失常。结论:在ARP期间施加电流可安全地显著增强心肌收缩力,提示CCM刺激是一种新型有效的收缩力调制方法。