Biosens Bioelectron. 2010 Jul 15;25(11):2425-30. doi: 10.1016/j.bios.2010.03.034. Epub 2010 Apr 3.
An improved telemetric system based on either differential pulse voltammetry (DPV) or direct current amperometry (DCA) "diffused" via a single-way infrared (IR) transmission channel is introduced. The structural design of this system is very small and weighting only few grams but above all and unlike similar pre-existing instruments based on IR transmission, the present system works on a single-way communication, thus avoiding problems related to cross-talking between two-way channels that ends in altered functionality of the measuring electrode system. In addition, it is immune from electromagnetic interferences that disturb radio-frequency transmission. The system appeared already sensitive to both dopamine (DA) and serotonin (5-HT) using DCA and DPV either in vitro as well as in vivo. In particular, parallel in vivo experiments within anaesthetised rats prepared for classical wire-connected DCA or for wireless DCA resulted in superimposing data when the 5-HT system in the frontal cortex of anaesthetised rats was challenged with fluoxetine. In the present study, we have further refined and then successfully extended the wireless methodology to conscious freely moving rats and further verified the chemical nature of the DA and 5-HT current related signals recorded with wireless DCA in prefrontal cortex using Nafion coated micro-biosensors via dedicated pharmacological treatments.
引入了一种基于差分脉冲伏安法(DPV)或直流安培法(DCA)的改良遥测系统,通过单向红外(IR)传输通道“扩散”。该系统的结构设计非常小巧,重量仅几克,但与基于 IR 传输的类似现有仪器不同,该系统采用单向通信,从而避免了双向通道之间串扰导致测量电极系统功能改变的问题。此外,它还能免受干扰射频传输的电磁干扰。该系统使用 DCA 和 DPV 已经对多巴胺(DA)和 5-羟色胺(5-HT)具有敏感性,无论是在体外还是体内。特别是,在为经典有线 DCA 或无线 DCA 准备的麻醉大鼠体内进行平行实验时,当麻醉大鼠额皮质中的 5-HT 系统受到氟西汀挑战时,数据会叠加。在本研究中,我们进一步改进了无线方法,然后成功地将其扩展到了清醒自由活动的大鼠,并进一步使用涂有 Nafion 的微生物传感器通过专门的药理学处理,验证了在额前皮质中使用无线 DCA 记录的与 DA 和 5-HT 电流相关信号的化学性质。