J Magn Reson Imaging. 2010 May;31(5):1277-81. doi: 10.1002/jmri.22147.
To test a newly developed fat suppression magnetic resonance imaging (MRI) prepulse that synergistically uses the principles of fat suppression via inversion recovery (STIR) and spectral fat saturation (CHESS), relative to pure CHESS and STIR. This new technique is termed dual fat suppression (Dual-FS).
To determine if Dual-FS could be chemically specific for fat, the phantom consisted of the fat-mimicking NiCl(2) aqueous solution, porcine fat, porcine muscle, and water was imaged with the three fat-suppression techniques. For Dual-FS and STIR, several inversion times were used. Signal intensities of each image obtained with each technique were compared. To determine if Dual-FS could be robust to magnetic field inhomogeneities, the phantom consisting of different NiCl(2) aqueous solutions, porcine fat, porcine muscle, and water was imaged with Dual-FS and CHESS at the several off-resonance frequencies. To compare fat suppression efficiency in vivo, 10 volunteer subjects were also imaged with the three fat-suppression techniques.
Dual-FS could suppress fat sufficiently within the inversion time of 110-140 msec, thus enabling differentiation between fat and fat-mimicking aqueous structures. Dual-FS was as robust to magnetic field inhomogeneities as STIR and less vulnerable than CHESS. The same results for fat suppression were obtained in volunteers.
The Dual-FS-STIR-CHESS is an alternative and promising fat suppression technique for turbo spin echo MRI.
测试一种新开发的磁共振成像(MRI)预脉冲脂肪抑制技术,该技术协同使用反转恢复(STIR)和频谱脂肪饱和(CHESS)的脂肪抑制原理,与纯 CHESS 和 STIR 相比。这项新技术被称为双重脂肪抑制(Dual-FS)。
为了确定 Dual-FS 是否可以对脂肪进行化学特异性抑制,该体模由模仿脂肪的 NiCl2 水溶液、猪脂肪、猪肌肉和水组成,使用三种脂肪抑制技术对其进行成像。对于 Dual-FS 和 STIR,使用了几种反转时间。比较了每种技术获得的每个图像的信号强度。为了确定 Dual-FS 是否可以对磁场不均匀性具有鲁棒性,使用不同浓度的 NiCl2 水溶液、猪脂肪、猪肌肉和水组成的体模,在几个失谐频率下使用 Dual-FS 和 CHESS 进行成像。为了比较体内脂肪抑制效率,还对 10 名志愿者进行了三种脂肪抑制技术的成像。
Dual-FS 可以在 110-140msec 的反转时间内充分抑制脂肪,从而能够区分脂肪和模仿脂肪的水结构。Dual-FS 对磁场不均匀性的鲁棒性与 STIR 相同,并且比 CHESS 更不易受影响。在志愿者中也获得了相同的脂肪抑制效果。
Dual-FS-STIR-CHESS 是一种用于涡轮自旋回波 MRI 的替代且有前途的脂肪抑制技术。