INTRODUCTION

The CO2 laser has a long history in both experimental and clinical neurosurgery. However, its use over the past decade has been limited by its cumbersome design and bulky set-up of the micromanipulator. These limitations are amplified when it is used with the operating microscope. These restrictions are addressed by the Omniguide fiber, which delivers the beam through flexible hollow-core photonic bandgap mirror fibers and allows the laser to be wielded like any other surgical instrument.

METHODS

The attending neurosurgeon prospectively assessed the usefulness of the laser in its first 45 consecutive uses at our institution based on a scale of 1 to 5.

RESULTS

The series included 11 cavernous malformations, 14 meningiomas, 7 ependymomas, 3 metastases, 3 astrocytomas, and 7 miscellaneous lesions. The laser was set up 91 times and used in 45 cases. The Omniguide fiber failed 5 times. No adverse events involving patients or staff were associated with laser use. The mean utility score was 3.7 +/- 0.8 (range, 2-5). The laser was most helpful in debulking fibrous lesions too tough for ultrasonic aspiration and lesions adherent to delicate neurovascular structures. The laser was not helpful with highly vascular tumors.

CONCLUSION

In our early experience, the Omniguide laser was very helpful in selected cases in resecting specific types of lesions without complications; we have added the device to our neurosurgical armamentarium.

摘要

简介

CO2 激光在实验和临床神经外科中都有着悠久的历史。然而,在过去十年中,由于其繁琐的设计和庞大的显微镜操纵器设置,其使用受到限制。当与手术显微镜一起使用时,这些限制会被放大。这些限制可以通过传递光束的光纤来解决,这种光纤可以通过柔性中空芯光子带隙镜纤维来传输,从而使激光能够像任何其他手术器械一样使用。

方法

主治神经外科医生根据 1 到 5 的评分标准,在我们机构对该激光的前 45 次连续使用情况进行了前瞻性评估。

结果

该系列包括 11 例海绵状畸形、14 例脑膜瘤、7 例室管膜瘤、3 例转移瘤、3 例星形细胞瘤和 7 例其他病变。该激光共设置了 91 次,在 45 例病例中使用。Omniguide 光纤出现故障 5 次。激光使用没有发生任何涉及患者或医护人员的不良事件。平均使用评分 3.7 +/- 0.8(范围 2-5)。该激光在切除对超声吸引过于坚韧的纤维性病变和与精细的神经血管结构粘连的病变方面非常有帮助。但该激光对高度血管化的肿瘤没有帮助。

结论

在我们的早期经验中,Omniguide 激光在切除特定类型的病变方面非常有帮助,没有并发症;我们已经将该设备添加到我们的神经外科武器库中。