Department of Ophthalmology, Mayo Clinic, Rochester, Minnesota, USA.
Invest Ophthalmol Vis Sci. 2010 Nov;51(11):5610-6. doi: 10.1167/iovs.10-5614. Epub 2010 Jun 10.
Corneal stromal haze cannot be compared in longitudinal studies or across laboratories without standardization. In this study, a method was devised of standardizing image brightness in confocal microscopy of the cornea.
Thirty-six normal corneas of 18 untreated volunteers and 35 corneas of 18 patients 3 years after LASIK were examined by slit scanning confocal microscopy. The mean image intensity of each frame was adjusted for daily variations in sensitivity of the microscope by using scans through a solution of Amco Clear (GSF Chemicals, Columbus, OH). Adjusted image intensities were expressed in scatter units (SU), representing the concentration of Amco Clear that produced the same intensity as the image. The intensity from the stroma in corneas after LASIK was compared to that in untreated corneas by using generalized estimating equation models.
In the untreated corneas, image brightness was 1079 ± 242 and 758 ± 142 SU in the anterior and mid stroma, respectively. Three years after LASIK, image intensity in the flap was 740 ± 186 SU, approximately 30% lower than in corresponding stroma of the untreated corneas (P < 0.001). At mid stroma, brightness was 715 ± 117 SU after LASIK, and was not significantly different from brightness in untreated corneas (P = 0.26).
Clinical confocal microscopy provides a high-resolution measurement of corneal haze, and Amco Clear provides a means of standardizing these measurements. This method can detect subtle decreases in haze in the corneal flap 3 years after LASIK and could be used to examine changes in haze after lamellar keratoplasty.
如果不对角膜基质混浊进行标准化,就无法在纵向研究或跨实验室进行比较。本研究旨在设计一种标准化角膜共焦显微镜图像亮度的方法。
对 18 名未经治疗的志愿者的 36 只正常角膜和 LASIK 术后 3 年的 18 名患者的 35 只角膜进行了裂隙扫描共焦显微镜检查。通过扫描 Amco Clear(GSF Chemicals,Columbus,OH)溶液,调整每个帧的平均图像强度,以适应显微镜每日灵敏度的变化。调整后的图像强度以散射单位(SU)表示,代表产生与图像相同强度的 Amco Clear 的浓度。使用广义估计方程模型比较 LASIK 后角膜的基质强度与未经治疗的角膜的强度。
在未经治疗的角膜中,前基质和中基质的图像亮度分别为 1079 ± 242 和 758 ± 142 SU。LASIK 术后 3 年,瓣层的图像强度为 740 ± 186 SU,比未经治疗的角膜相应基质低约 30%(P < 0.001)。在中基质,LASIK 后亮度为 715 ± 117 SU,与未经治疗的角膜无显著差异(P = 0.26)。