Department of Radiology, Yamaguchi University Graduate School of Medicine, 1-1-1 Minamikogushi, Ube, Yamaguchi, 755-8505, Japan.
Neuroradiology. 2011 Apr;53(4):255-60. doi: 10.1007/s00234-010-0730-5. Epub 2010 Jun 29.
There has been concern regarding the usefulness of diffusion-weighted imaging (DWI) to evaluate the ischemic lesions associated with carotid artery stent placement (CAS). Some small lesions may be detected not by standard DWI but by thin-slice DWI alone, since most of the cerebral lesions are very small in size and clinically silent. The purpose of this study is to compare the detectability of the small ischemic lesions after CAS by standard and thin-slice DWI.
Both standard DWI with slice thickness of 6 mm and thin-slice DWI with slice thickness of 2 mm were obtained at the same MR examination within 2 to 7 days after 20 procedures of CAS in 17 patients. Number and measured diameter size of the detected lesions on both DWI were compared.
All CAS procedures in 17 patients were successfully completed. The focal ischemic lesions were detected in 14 of 20 on thin-slice DWI and seven examinations on standard DWI. The total numbers of hyperintense lesions were 31 on thin-slice DWI and ten on standard DWI (p < 0.001). The sizes of these ten lesions on thin-slice DWI were larger than those of standard DWI, and the mean size of the thin-slice DWI and that of standard DWI were significantly different (p < 0.005).
Thin-slice DWI was able to detect small cortical lesions better than standard DWI. Thin-slice DWI may be useful to evaluate small silent ischemic lesions after CAS.
人们一直担心扩散加权成像(DWI)在评估颈动脉支架置入(CAS)相关缺血性病变方面的作用。一些小病变可能无法通过标准 DWI 检测到,而仅通过薄层 DWI 检测到,因为大多数脑病变的体积非常小且临床表现为无症状。本研究旨在比较标准 DWI 和薄层 DWI 对 CAS 后小缺血性病变的检测能力。
在 17 例患者的 20 次 CAS 后 2 至 7 天内的同一次磁共振检查中,同时获得标准 DWI(层厚 6mm)和薄层 DWI(层厚 2mm)。比较两种 DWI 上检测到的病变数量和测量直径大小。
17 例患者的所有 CAS 操作均顺利完成。在 20 次薄层 DWI 检查中有 14 次和 7 次标准 DWI 检查中检测到局灶性缺血性病变。在薄层 DWI 上共检测到 31 个高信号病变,在标准 DWI 上检测到 10 个(p<0.001)。薄层 DWI 上这 10 个病变的大小大于标准 DWI,且两者的平均大小差异具有统计学意义(p<0.005)。
薄层 DWI 比标准 DWI 更能检测到小皮质病变。薄层 DWI 可能有助于评估 CAS 后小的无症状性缺血性病变。