Division of Cardiology, VA Pittsburgh Healthcare System and the University of Pittsburgh Medical Center, Pittsburgh, Pennsylvania, USA.
Clin Cardiol. 2010 Aug;33(8):466-75. doi: 10.1002/clc.20803.
Late intracardiac lead perforation is defined as migration and perforation of an implanted lead after 1 month of cardiac electronic device implantation. It is an under-recognized complication with significant morbidity and mortality, particularly if not recognized early. Two patients with late perforation caused by passive-fixation leads are reported and the clinical features of their presentation and management are reviewed. We conducted a thorough review of the available English language literature pertaining to this complication to draw relevant conclusions regarding presentation, diagnosis, and management. Early recognition of this complication is important as the indications for and numbers of patients who receive cardiac implantable electronic devices continue to expand.
心脏内导线迟发性穿孔是指在心脏电子设备植入 1 个月后,植入的导线发生迁移和穿孔。这是一种认识不足的并发症,具有较高的发病率和死亡率,尤其是如果不能早期发现。本文报告了 2 例由被动固定导线引起的迟发性穿孔患者,并回顾了其临床表现和处理方法。我们对相关英文文献进行了全面回顾,以得出有关该并发症的临床表现、诊断和处理的相关结论。由于植入式心脏电子设备的适应证和患者数量不断增加,因此早期识别这种并发症非常重要。