Modulation transfer function measurement technique for small-pixel detectors.

作者信息

Marchywka M, Socker D G

出版信息

Appl Opt. 1992 Dec 1;31(34):7198-213. doi: 10.1364/AO.31.007198.

Abstract

A modulation transfer function (MTF) measurement technique suitable for large-format, small-pixel detector characterization has been investigated. A volume interference grating is used as a test image instead of the bar or sine wave target images normally used. This technique permits a high-contrast, large-area, sinusoidal intensity distribution to illuminate the device being tested, avoiding the need to deconvolve raw data with imaging system characteristics. A high-confidence MTF result at spatial frequencies near 200 cycles/mm is obtained. We present results at several visible light wavelengths with a 6.8-microm-pixel CCD. Pixel response functions are derived from the MTF results.