Suppr超能文献

植入时左心室刺激特征与心脏再同步治疗 6 个月后的超声心动图反应之间的关系。

Relationship between left ventricular stimulation characteristics at implantation and echocardiographic response after 6 months of cardiac resynchronization therapy.

机构信息

Hopital Cardiologique HautLeveque, University Bordeaux 2, CHU of Bordeaux, Bordeaux-Pessac, France.

出版信息

Europace. 2010 Dec;12(12):1757-61. doi: 10.1093/europace/euq301. Epub 2010 Sep 8.

Abstract

AIMS

Although the electrical stimulation of an ischaemic tissue adversely affects the left ventricular (LV) systolic function, the optimal stimulation site in patients with non-ischaemic cardiomyopathy has not been systematically studied. We hypothesized that the local stimulation characteristics at the time of device implantation predict the response to cardiac resynchronization therapy (CRT).

METHODS AND RESULTS

We measured the impedance, sensing, and capture threshold of a bipolar LV lead in 138 patients with non-ischaemic cardiomyopathy undergoing first implantation of CRT device for drug refractory heart failure. All patients underwent echocardiography at baseline and at 6 months post-implantation. An absence of favourable response to CRT was defined as <15% decrease in echocardiographic LV end-systolic volume (LVESV) at 6 months. Echocardiographic response to CRT was observed in 70% of patients. The LV lead measurements predicted neither the optimal stimulation site nor the response to CRT. Left ventricular capture threshold (1.50 ± 1.1 vs. 1.32 ± 0.8 V) and impedance (725 ± 287 vs. 720 ± 261 Ω) were similar between the responders and the non-responders. Independent of baseline LV ejection fraction or ESV, the LV R-wave amplitude at implantation was significantly higher (P = 0.0038) in responders (12.7 ± 5.2 mV) than in non-responders (9.7 ± 6.3 mV), with an area under the receiver operating characteristic curve of 0.7.

CONCLUSION

Response to CRT, as determined by decrease in LVESV at 6 months, was associated with significantly higher LV R-wave amplitude at the time of device implantation.

摘要

目的

尽管电刺激缺血组织会对左心室(LV)收缩功能产生不利影响,但对于非缺血性心肌病患者,最佳刺激部位尚未得到系统研究。我们假设,在器械植入时的局部刺激特征可预测心脏再同步治疗(CRT)的反应。

方法和结果

我们测量了 138 例接受 CRT 器械首次植入以治疗药物难治性心力衰竭的非缺血性心肌病患者的双极 LV 导联的阻抗、感知和捕获阈值。所有患者均在基线和植入后 6 个月进行了超声心动图检查。将 CRT 无明显反应定义为 6 个月时超声心动图 LV 收缩末期容积(LVESV)减少<15%。70%的患者对 CRT 有反应。LV 导联测量值既不能预测最佳刺激部位,也不能预测 CRT 的反应。LV 捕获阈值(1.50±1.1 比 1.32±0.8 V)和阻抗(725±287 比 720±261 Ω)在有反应者和无反应者之间相似。独立于基线 LV 射血分数或 ESV,植入时的 LV R 波振幅在有反应者(12.7±5.2 mV)中明显高于无反应者(9.7±6.3 mV),受试者工作特征曲线下面积为 0.7。

结论

以 6 个月时 LVESV 减少来确定 CRT 反应与器械植入时 LV R 波振幅显著升高相关。

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验